
Command reference
R&S
®
ZND
684
User Manual 1173.9557.02 ─ 61
Example:
*RST
SENS:FREQ:STAR 1GHz
SENS:FREQ:STOP 3GHz
CALC:LIM:CONT 1GHz, 2GHz, 2GHz, 3GHz
CALC:LIM:SEGM1:FORM '22-(20/25.78)*StimVal/1e9'
CALC:LIM:SEGM1:FORM:STAT ON
CALC:LIM:SEGM2:FORM '15-(6/25.78)*StimVal/1e9'
CALC:LIM:SEGM2:TYPE LMIN
CALC:LIM:DISP ON
CALC:LIM:STAT ON
Manual operation:
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula:STATe
<State>
Defines whether a custom formula is used for limit line segment <Seg> of the related
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
If you want to use a custom formula for a segment, use
CALCulate<Chn>:LIMit:CONTrol[:DATA]
Parameters:
<State>
OFF (0)
The line segment is a straight line, connecting the endpoints
specified in the segment definition.
ON (1)
The line segment is defined by a custom formula. Use
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
ify the formula.
*RST:
OFF
Example:
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
Manual operation:
CALCulate<Chn>:LIMit:SEGMent<Seg>:INTerpol
<LimLineInterpolation>
"Interpolation" determines whether the limit line segment is interpolated linearly or log-
arithmically. The latter may be more suitable for logarithmic sweeps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<LimLineInterpolation>LOGarithmic | LINear
SCPI command reference
www.allice.de
Allice Messtechnik GmbH