GUI reference
R&S
®
ZND
413
User Manual 1173.9557.02 ─ 61
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference
.
Defaults are ±0.2
dB for magnitude and ±2° for phase deviations.
Remote command:
Preferences
Allows you to configure the behavior of the validation diagrams and traces.
Unchecking "Auto Scale" prevents the traces from scaling as the measurement and/or
standard changes. Unchecking "Average On" turns off averaging.
Basic Mode
Switches to
5.11.2
Cal Devices tab
Provides access to all functions for calibration kit management and cal unit characteri-
zation.
Cal softtool