After successful measurement receiver calibration, the "b<n>(P<m>)" trace must be
close to the "a<m>(P<m>)" trace. Due to the previous power calibration, both traces
must be close to the cal power.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:
RESult?
Port Overview
Shows all source ports together with the possible power calibrations.
For each analyzer port P1 ... PN, a "Ref. Receiver", a "Meas. Receiver" and a "Source
Flatness" calibration can be performed.
Note:
For a unidirectional R&S
ZND (see
on page 80) P2 cannot serve as source port. Hence only a
calibration of the measurement receiver can be performed at P2.
Select a "Ref. Receiver", "Meas. Receiver", or "Source Flatness" symbol to start the
related calibration. See
"Power Cal dialog – Ref. Receiver"
"Power Cal dialog – Source Flatness"
A green checkmark indicates that the calibration data has been acquired successfully.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled when a new set of power calibration data has been acquired. The button
applies all available power calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
The power calibration state is indicated in the trace list, see
on page 178. Use the functions in the
on page 434 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
[SENSe<Chn>:]CORRection:PSTate?