GUI reference
R&S
®
ZND
409
User Manual 1173.9557.02 ─ 61
Power Cal dialog – Meas. Receiver
In "Meas. Receiver" mode, the "Power Cal" dialog guides you through the calibration of
a measurement receiver. To ensure accurate source signal powers, a source power
calibration must be performed before the measurement receiver calibration.
Background information
Chapter 4.5.6.2, "Measurement receiver calibration"
Access:
> "Meas. Receiver"
left: = source and meas. port identical
right = source and meas. port different
Port Overview
The dialog shows all receiver ports of the network analyzer. The selected port is dis-
played with the current cal power settings (see
Chapter 5.11.3.2, "Modify Cal Power
on page 430); moreover, a circuit diagram visualizes the purpose of the mea-
surement receiver calibration.
"Source Port" defines the type of measurement receiver calibration:
●
If the source port is equal to the calibrated port, the measurement receiver is cali-
brated by the wave that is reflected back by a connected Open or Short standard.
Connect the Open or Short standard to the calibrated port; no additional external
test setup is required.