Command reference
R&S
®
ZND
1037
User Manual 1173.9557.02 ─ 61
Manual operation:
SOURce:POWer:CORRection:PPOWer
<RecCalPower>
Defines the source power the R&S
ZND uses to perform the first calibration sweep of
the source power calibration ("Reference Receiver Cal Power"). This predefined power
value is ignored if the R&S
ZND is set to use the calculated port power result
(
SOURce:POWer:CORRection:PSELect
CPOW
).
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
"Reference Receiver Cal Power"
SOURce:POWer:CORRection:PSELect
<CalPower>
Defines how to define the source power the R&S
ZND uses to perform the first calibra-
tion sweep of the source power calibration ("Reference Receiver Cal Power").
Parameters:
<CalPower>
CPOWer | PPOWer
CPOWer
Calculated port power result
PPOWer
Predefined port power value, see
*RST:
CPOWer
Example:
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for port number
<PhyPt>
. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (
SOURce<Ch>:POWer:CORRection[:ACQuire]
).
The query always returns a result.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number