Command reference
R&S
®
ZND
1001
User Manual 1173.9557.02 ─ 61
[SENSe<Ch>:]SEGMent<Seg>:DELete[:DUMMy]
Deletes the specified (single) sweep segment.
deletes all segments in the channel.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:DEL
Delete the created segment.
Manual operation:
"Add / Insert / Delete / Delete All"
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer?
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?
These commands return the center frequency and the span (width) of sweep segment
no.
<Seg>
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:CENT? SPAN?
Query the center frequency and span of the created segment.
The response is
1250000;500000
.
Usage:
Query only
Manual operation:
Note:
The frequency range of the sweep segment can be changed via
and
.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt
<StartFreq>
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP
<StopFreq>
These commands define the start and stop frequency of sweep segment no.
<Seg>
.
The sweep segments must be within the frequency range of the R&S
ZND; see
.
Suffix:
<Ch>
.
Channel number
SCPI command reference