DT164-W User Guide
AP11870 Issue 1
Other options
The DT164-W will fit
a 12x12” NEMA Type 1
/ IP10 box or
equivalent. A custom mounting frame might be required.
Permanent installation with the gland module
⚠
Replacement of the IEC AC inlet module and mains
wiring for permanent installation must be performed by a
qualified electrician and carried out in accordance with
local wiring regulations.
To replace the IEC module:
•
Remove the 4 screws using a Torx T8 driver.
•
Pull out the module and unplug the harness at the bottom of the
assembly.
•
Fit a suitable cable gland or grommet appropriate to the cable wire
size to the gland module plate.
•
Run the mains cable through the gland. Minimum cable size 0.75mm
2
or 20 AWG.
•
Wire the mains cable to the terminal connector on the gland module,
paying attention to polarity and earth grounding as indicated on the
module.
•
Securely tighten the cable gland around the cable.
•
Re-plug the internal harness into the bottom of the gland module.
•
Secure the module to the unit with the 4x T8 screws.
Terminal for
mains connection
Harness for
internal
connection