Instructions for Use for KardiaMobile Card (AC-021) 

 

 

 

 

 

 

 

 

 

 

 

Page 9 of 17 

 

Instant Analysis 

Description 

Additional information 

Bradycardia 

Your heart rate is less than 50 beats per 
minute, which is slower than normal for 
most people.  

Kardia does not check for heart 
attack. If you believe you are having 
a medical emergency, call 
emergency services. Do not change 
your medication without talking to 
your doctor. 

Normal 

 

No rhythm abnormalities detected in 
your EKG.   
 
 

Kardia does not check for heart 
attack. If you believe you are having 
a medical emergency, call 
emergency services. Do not change 
your medication without talking to 
your doctor. 

Tachycardia

 

Your heart rate is faster than 100 beats 
per minute. This can be normal with 
stress or physical activity.  

Kardia does not check for heart 
attack. If you believe you are having 
a medical emergency, call 
emergency services. Do not change 
your medication without talking to 
your doctor. 

No Analysis

 

Your EKG recording is of insufficient 
duration. Instant Analysis is not able to 
provide an analysis on EKG recordings 
shorter than 30 seconds. 

Record a new EKG. Try to relax and 
hold still, rest your arms, or move to 
a quiet location that will allow for a 
full 30 second recording. 

Unclassified

 

Atrial fibrillation was not detected and 
your EKG does not fall under the 
algorithmic classifications of Normal, 
Bradycardia, or Tachycardia. This may 
be caused by other arrhythmias, 
unusually fast or slow heart rates, or 
poor quality recordings. 

Kardia does not check for heart 
attack. If you believe you are having 
a medical emergency, call 
emergency services. Do not change 
your medication without talking to 
your doctor. 

Unreadable

 

There is too much interference in this 
recording.  

Please re-record the EKG. Try to 
relax and hold still, rest your arms, 
or move to a quiet location or away 
from electronics and machinery. 

 

WARNING:

 After EKG analysis, the app may incorrectly identify ventricular flutter, ventricular bigeminy, and 

ventricular trigeminy heart conditions as unreadable. Please consult with your physician. 

 

NOTE: 

All historical EKGs and Instant Analysis results can be viewed, downloaded, and 

emailed from the “History” section of the Kardia app. 
 
EKG reports viewed at any magnification other than 100% may appear distorted and could lead 
to misdiagnosis. 
 

Summary of Contents for AC-021

Page 1: ...trademarks of AliveCor Inc in the United States and other jurisdictions Instructions for Use IFU for KardiaMobile Card AC 021 21LB01 01 December 2021 AliveCor Inc 189 N Bernardo Avenue Suite 100 Moun...

Page 2: ...ionality 6 Setting up your KardiaMobile Card device for the first time 7 Recording an EKG 7 EKG Analysis 8 Heart Rate 10 Environmental Specifications 10 Expected Service Life 10 Maintenance 10 Device...

Page 3: ...your heart rhythm is provided upon completion of your EKG recording See the EKG Analysis section for more details 4 KardiaMobile Card requires a compatible smartphone or tablet and the Kardia app a T...

Page 4: ...ny EKG result including normal You should notify your physician for possible changes in your health or if you are experiencing any concerning symptoms 2 Chest pain or pressure is a medical emergency H...

Page 5: ...making 12 The device has not been tested for and is not intended for pediatric use 13 Device contains a lithium manganese dioxide battery that is not removable or replaceable 14 Do not use the electr...

Page 6: ...ard System is intended for use by healthcare professionals patients with known or suspected heart conditions and health conscious individuals The device has not been tested and is not intended for ped...

Page 7: ...evice list at www alivecor com compatibility 3 Make sure Bluetooth is turned on in your smartphone or tablet settings 4 Launch the Kardia app and tap Create Account 5 Follow the on screen instructions...

Page 8: ...G recording is complete 8 The device will turn off automatically after use EKG Analysis Upon completion of your EKG recording the EKG is processed by AliveCor s Instant Analysis algorithms in the Kard...

Page 9: ...recordings shorter than 30 seconds Record a new EKG Try to relax and hold still rest your arms or move to a quiet location that will allow for a full 30 second recording Unclassified Atrial fibrillat...

Page 10: ...favorably compared to Lead I recordings of a standard FDA cleared 12 lead device Clinical equivalence of recordings from these Kardia devices and the 12 lead EKG device were also validated by board c...

Page 11: ...ces KardiaMobile Card has been tested and deemed in conformance with the relevant requirements in IEC 60601 1 2 2014 Class B for Electromagnetic Compatibility EMC FCC and Industry Canada IC Compliance...

Page 12: ...s open the KardiaMobile Card device is selected and is ready to pair Make sure the amber LED light turns on when you press the white power button located inside the A symbol on the right corner of the...

Page 13: ...recording Ensure that the Enhanced Filter is on Make sure Mains Filter is set appropriately for your geographical location This can be adjusted under the Kardia app Settings 3 On my EKG the recording...

Page 14: ...immunity KardiaMobile Card is intended for use in the electromagnetic environment specified below The customer or the user of KardiaMobile Card should assure that it is used in such an environment Im...

Page 15: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symb...

Page 16: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1...

Page 17: ...Symbol Interpretation Manufacturer Refer to instruction manual booklet Read instructions before use Do not dispose with household waste Temperature range Do not cut shred or attempt to destroy device...

Reviews: