AUF610 Series Operation Manual
OP610.1.1.5R1ENG
24
M43
Clear the zero point set by the user, and restore the zero point set by the manufacturer
M44
Set up a manual flow bias. Generally this value should be 0.
M45
Scale factor for the instrument. The default value is ‘1’.
M46
Network environment Identification Number. Any integer can be entered except
13(0DH, carriage return), 10 (0AH, line feeding), 42 (2AH), 38, 65535.
Every set of the instrument in a network environment should have a unique IDN.
Please refer to the chapter for communication.
M47
System lock-out to avoid inadvertent modification of the parameters
M48 Not
used
M49
Window for Network Communication Testing
M50
“Window to set up schedule for time based data saving. Selects items to be saved.
M51
Time setup for the data logger
M52
Data logging direction control. If ‘To RS-232’ is selected, all the data produced by data
logger will be transmitted out through the RS-232 interface
If ‘To buffer ‘is selected, the data will be stored into the built-in logger memory
Buffer transferring and buffer clearing
M53
Logger buffer viewer. It functions as a file editor. Use Dot, backspace UP and DN keys
to browse the buffer.
M54 Not
used
M55 Nod
used
M56 Not
used
M57 Not
used
M58 Not
used
M59 Not
used
M60
Settable 99-year calendar and Clock. Press ENT for modification. Use the dot key to
skip the digits that need no adjusting
M61
Display Version information and Electronic Serial Number (ESN) that are unique for
each AUF610 series flow mater.
M62
RS-232 setup. Baud rate & Parity – Adjustable from 75 to 115200 bps
M63 Not
used
M64 Not
used
M65 Not
used
M66 Not
used
M67
Input the frequency range for the frequency output. The biggest range is 0Hz-9999Hz.
M68
Enter a flow rate value that corresponds to lower frequency
M69
Enter a flow Rate value that corresponds to higher frequency
M70
LCD display backlight control. The entered value indicates how many seconds the
backlight will be on with each key stroke.
M71
LCD contrast control. The LCD will become darker when a small value is entered.
M72
Working timer. It can be cleared by pressing ENT key, and then select YES.
M73
Alarm #1 Low Flow Rate Threshold value. There are two available alarm indications in
the system. Either an Open Collector or a buzzer. Select in menu items M78 and
M77
M74
Enter the higher flow rate value that will trigger the #1 Alarm.