4 Installation
4.9
Calibrate - step 3 - Rotacheck Basic
Step 1
The Rotacheck Basic must be fully installed:
- Mechanically
- Electrically
The unit will flash green on power up and perform an internal
calibration after 5 sec.
The unit is ready to use when the green LED lights steadily.
Optional re-calibration
Push the blue button (N) on the sensor unit until you see a yellow
flash.
The Rotacheck Basic will automatically re-calibrate.
4512-0020
N
Step 2
(If unit has been opened)
Put the prism/top cover back on the base part by pushing it down
when the mark on the prism (J) and the open padlock (K) are
aligned. Then turn it clockwise towards the closed padlock (L) to
secure. Counterhold on the base part.
4512-0021
J
K
L
Operating LED and PLC feedback
Rotacheck Basic status
LED feedback
PLC feedback
Calibration
Green, flashing
No feedback
Idle
Green, steady
On
Hit
Yellow, 1 sec
On, 1 sec
Alarm*
Red, steady
On
* Unit failure or constant hit (cleaning device error)
19