background image

 

Troubleshooting 

 

WARNING DISPLAY

 “

Flow

”: If you see “Flow” in the display, the breath sample was not strong or 

long enough for measurement. Please try again and make sure the breath is strong and steady 
during sampling.  As soon as you start blowing into the device, you will hear a long beep indicating 
how long you should blow.

 

 

WARNING DISPLAY

 “

bAt

”: If you see “bAt” in the display, the battery power is low. Please replace 

the batteries with a new pair.  Make sure to properly align the batteries by looking at the +/- marks.

 

 

WARNING DISPLAY

 "

Err2/Err4

":

 If either of these codes appear in the display, the device is 

reporting a general problem with the alcohol sensor, or that the device has been physically 
damaged.  Please replace sensors or contact AK GlobalTech Corp. 

 

WARNING DISPLAY

 "

Er

r6

"

: The device is reporting that the sensor module is either 

improperly connected or missing altogether. 

It is very important when installing a 

sensor module that the 9-pin contacts are securely in position and that the central tab 
is locked in place. 

  

 

WARNING DISPLAY

 "

Err9

"

: Power source is too weak to complete operation.  Please replace 

batteries with a new pair. 

 

For the other troubles, please contact AK GlobalTech at 

[email protected]

, along with your 

error codes and photos if available.   

 

- Specifications - 

 

Alcohol Sensor  Advanced Electro-Chemical Fuel-Cell 

Detection Range  0.000 -0.400 g/dL (Accuracy to 0.02) 

Warm-up Time  5 seconds - 2 minutes 

Response Time  Within 10 seconds at 0.100%BAC 

Calibration  In lieu of traditional calibration, the sensor module must be calibrated every 

1,000 tests or at least once yearly, whichever comes first, in situations 
directly overseen or mandated by the DOT, users must perform calibration 
checks according to the manufacturer's QAP for the Revo TS200. 

Breath Sampling  Sample complete on audible click 

Sensor Accuracy  +/-0.005 up to 0.100 g/dL 

Display  4-Digit LCD 

Battery  Two 1.5V AAA Alkaline (included) 

Weight  3.0 oz 

Dimensions  4.0 x 2.0 x 0.7 in 

Casing  Gray ABS Plastic 

Operating Temperature  32-104 degrees Fahrenheit or  

0-40 degrees Celcius 

Warranty  One (1) Year (Refer to "Warranty" section 

Summary of Contents for Revo TS200

Page 1: ...1 Revo Model TS200 Revo Model TS200 Operator s Manual ...

Page 2: ...2 ...

Page 3: ...able design The device also incorporates patented pre calibrated replaceable sensor module technology which means that unlike other typical breathalyzers it does not have to be mailed back for calibration service To ensure accuracy users of the Model TS200 in situations directly overseen or mandated by the US Department of Transportation DOT are required to conduct accuracy checks before and after...

Page 4: ...Cover Removal Remove the rear cover then remove the batteries Use your finger to hold down the small tab highlighted to the left and pull the sensor module straight down You can then simply lift the module up from the device To install a new sensor module place the portion of the sensor module highlighted to the left flush against the section of the slot also highlighted You can then simply slide ...

Page 5: ... the mouthpieces slot 6 Blow sharply into the device until you hear the first audible click and continue to steadily blow until you hear the second audible click which signals that sampling is complete You should continue to blow more than 3 seconds until you hear the second audible clicks 7 After a successful sample the device will analyze the sample for a few seconds you will see the digit 0 mov...

Page 6: ... this recorded date or approximately 1000 tests whichever comes first Warnings and Error Codes Warnings and Precautions WAIT AT LEAST 20 MINUTES AFTER EATING OR DRINKING BEFORE TESTING A breathalyzer examines lung air samples not mouth air to calculate alcohol content Any contaminants in the mouth including alcohol food smoke or even excess saliva may interfere with testing This includes alcohol b...

Page 7: ...ontacts are securely in position and that the central tab is locked in place WARNING DISPLAY Err9 Power source is too weak to complete operation Please replace batteries with a new pair For the other troubles please contact AK GlobalTech at support alcomate net along with your error codes and photos if available Specifications Alcohol Sensor Advanced Electro Chemical Fuel Cell Detection Range 0 00...

Page 8: ...ol including whether the test subject is intoxicated under any definition of that word AK GlobalTech expressly disclaims any liability for direct indirect incidental special or consequential damages of any nature under any legal theory Any act or failure to act based on a reading from this product shall be at the test subject s or user s own risk AK GlobalTech assumes no responsibility for consequ...

Page 9: ... reseller agent or employee is authorized to make any modification extension or addition to this warranty If any term is held to be illegal or unenforceable the legality or enforceability of the remaining terms shall not be affected or impaired EXCEPT AS PROVIDED IN THIS WARRANTY AND TO THE MAXIMUM EXTENT PERMITTED BY LAW AK GLOBALTECH IS NOT RESPONSIBLE FOR DIRECT SPEICAL INCIDENTAL OR CONSEQUENT...

Reviews: