5. Troubleshooting
B1924-05
Symptom
Cause
Remedy
Mechanics
(1) The image is
not focused while
using high power
objective.
The cover glass faces down.
Put the cover glass to face up.
The cover glass is not standard.
Use a standard cover glass
with thickness 0.17mm.
(2) The objective
touches the cover
glass while
turning the
nosepiece.
The cover glass faces down.
Put the cover glass to face up.
The cover glass is not standard.
Use a standard cover glass
with thickness 0.17mm.
(3)Can not move
the slide moothly.
The slide is not fixed correctly.
Adjust it correctly.
The movable specimen holder is not
fixed properly.
Tighten it.
Electrics
(1)The bulb does
not work
The battery is lower.
Recharging
Digital
(1)No video
USB connected incorrectly
Connect it in right way
Driver installed incorrectly
Install the driver correctly (See
readme.txt under the root dir)
(2) The video is
too light or too
dark.
Condenser adjusted improper
Adjust the condenser
Iris diaphragm adjusted improper
Adjust the Iris diaphragm
Brightness adjust incorrectly
Adjust the Brightness
(3) Unclear video
image.
Focusing incorrectly
Adjust focus