3) APR
a) AJEX240H is provided 9 APR position select key and 3 APR size key for
storage of exposure conditions which is used frequently.
☞ Refer to '3-5. Overview of operating elements' on the page 14.
b) If select APR position key or APR size key, change NORMAL MODE to APR
Mode and displayed exposure condition which is stored.
Select desired APR position key after selection of APR size key or select
inversely.
c) Return to NORMAL MODE
(a) If press again the APR position key or APR size key which is selected,
return to NORMAL MODE.
(b) If press the kV △,▽ key or mAs △,▽ key, return to NORMAL MODE.
d) Change and storage of exposure conditions for APR MODE.
(a) If press the SET key during 3 sec. when APR MODE is selected,
short audible buzzer sound is generated. SET key indicator glows green.
(b) Select APR position and APR size which will be changed.
(c) Press the kV △,▽ switch and mAs △,▽ switch for setting of desired
exposure conditions.
(d) Press the selected APR position key during 3 sec. when selection of
exposure conditions is finished.
After short audible buzzer sound, new exposure conditions are stored
in the memory appropriated to current APR position and size.
(e) If you want to change stored exposure conditions of other APR position
and size, repeat b) ~ d).
Otherwise, press the SET key during 3 sec. until short audible buzzer
sound is generated.
Then return to APR MODE after lights-out of SET key indicator.
-------------------------------------------------------------
AJEX240H / Operating instructions
-22-
AXR-240H-MN01 (E) / Rev. 03