7.0 Technical Data
7.1
SPECIFICATIONS
Regalia
Performance Data
1
Product Flow
2
Product Pressure
21 SCFH (0.55 Nm
3
/hr or 10 LPM)
1
7 psig (48 kPa or 0.48 barg)
1
Product Concentration
90% +/- 3%
Product Dew Point
-100°F (-73°C)
Dimensions (W x D x H)
(nominal)
14.7 x 19.5 x 26.5 in (37 x 50 x 67 cm)
Weight
57 lb (26 kg)
Physical Connections
Product Gas Outlet
DISS fitting
Storage Temperature
-4ºF (-20ºC) to 140ºF (60ºC)
Ambient Operating Conditions
Locate the oxygen concentrator system in a well-
ventilated area that is protected from weather elements
and remains between 50ºF (10ºC) and 104ºF (40ºC)
inside room.
Barometric Pressure Range: 28-31 of Hg
Ambient Oxygen Concentration: 20% minimum
ISO Class Reference
IEC Class II, Type BF Applied Part, Continuous
Operation
Control Power Requirements
(Single Phase)
115 V ~ ±10%, 60 Hz, 5.0 A
220 – 240 V ~ ±10%, 50 Hz, 2.4 A
Typical Power Consumption
600 W
Oxygen Concentration Indicator
Green Light = Normal Operation
Yellow Light Flashing = Below Normal Operation (85%*)
Red Light Flashing = Abnormal Operation (70%**)
Audible Alarm Indicators
Power Interruption
Power Surge (simulated circuit breaker)
Compressor Malfunction
Excessive Internal Temperature
Outlet Pressure
[35-36 psig (241.2-248.1 kPa) relief valve]
Low Oxygen Output** <70% oxygen
Back-up Alarm Power
9V Battery
Filters
Cabinet Air Inlet, Compressor Inlet and HEPA
Back pressure Effect on Flow
Indicator (nominal)
0 psig (0 kPa): 10.0 LPM
1 psig (7 kPa): 10.0 LPM
OSCI Operation*
“Temperature: 50º F to 104º F (10º C to 40º C)
Independent of Atmospheric Pressure. 85 +/-3% and 70
+/-5% oxygen concentration measurement accuracy”
*Where equipped. Subject to ±3% system concentration accuracy, and for backpressures up to 7 kPa.
**Where equipped. Subject to ±5% system concentration accuracy, and for backpressures up to 7 kPa.
1
Outlet pressure may decline to 5 psig (34.5 kPa) nominal at 10,000 ft altitude. The system performance may
decline above 10,000 ft (3,048 m).
2
SCF (Standard cubic foot) gas measured at 1 atmosphere and 70°F / Nm
3
(Normal cubic meter) gas measured
at 1 atmosphere and 0°C / LPM (Liters per minute) gas measured at 1 atmosphere and 21°C
Table 7.5:
Regalia Specifications