ImPulse
™
Elite Patient Manual
Keep the ImPulse
Elite clean and free from moisture, dust, and extreme
temperatures. To help prevent possible damage to the unit, keep ImPulse
Elite
in its carrying case.
Clean the unit periodically by wiping with a lint-free cloth. Pay special attention
to the oxygen outlet for the cannula connection to make sure it remains free of
dust, water, etc. Do not expose or submerge the ImPulse
Elite device in water.
We recommend using a one-part bleach to 99 parts water solution or using
Fantastik
®
spray cleaner as a safe cleaner for the ImPulse Elite’s outer
housing.
A list of undesirable chemical cleaning agents includes, but is not limited to
the following, according to the plastic manufacturer: alcohol and alcohol-
based products (isopropyl alcohol), concentrated chlorine-based products
(Ethylene Chloride), and oil-based products (Pine-Sol, Lestoil). These are
NOT to be used to clean the plastic housing on the ImPulse Elite OCD. These
chemicals and the cleaners that contain them will crystallize the plastic and
break down the plastic’s molecular properties.
Replace the disposable cannula periodically following normal usage.
Additional supplies are available from your Equipment Provider or AirSep.
Return empty oxygen cylinders to your supplier for replacement. Remove the
unit from the empty cylinder before returning the cylinder to your supplier.
Note
Do not open the electronic compartment to service the device. Open only the battery
compartment when necessary. To prevent a voided AirSep warranty, follow all
manufacturer’s instructions.
20
Proper Care & Maintenance