![AirClean systems AC-DS-03 Instruction Manual Download Page 7](http://html1.mh-extra.com/html/airclean-systems/ac-ds-03/ac-ds-03_instruction-manual_2876280007.webp)
AC-DS-03 Automated Deactivation System Manual Ver. A03
3
Securing your AC-DS-03
(Tools needed: #2 Phillips head screwdriver, pencil, electric drill)
1.
There must be a minimum of 4” of clearance on the left side of the unit to allow for access to
the power receptacle and fuse.
2.
Place unit within 8 feet of drain connection.
3.
Remove screws from either side of the AC-DS-03 and place the slotted wall bracket over the
screw hole on the AC-DS-03 and install screws loosely.
4.
Position unit against wall for wall bracket installation. Push bracket against wall and mark
position of bracket hole. See Figure 1.
5.
Repeat on other side of unit.
6.
Install the two self-drilling wall anchors at these points until they are flush with the wall. Place
bracket over anchor and install Phillips head screws (do not over-tighten). Now tighten
7.
For masonry walls use suitable lead or plastic anchors and bolts. These anchors are not
supplied.
8.
The AC-DS-03 unit has 4 adjustable feet which can be screwed up or down to level the unit.
Figure 1: AC-DS-03 brackets against wall.
Figure 2: Drain fitting