Running

When the pump is switched on the timer will display the time elapsed in 

minutes. 

When the pump stops, the timer will display the elapsed time. If the pump is 

restarted, the timer will begin to count from the time at which it stopped. 

Alternatively the display can be zeroed using the reset button (4).

Charging

Each pump comes with a voltage detecting battery charger. This is preset to 

give 500mA charge rate for sufficient time to recharge the battery pack. 
At this rate the battery will be fully recharged from flat in approx 5 hours.
When the charger is plugged into the charging socket the red LED will be 

illuminated. 
When the battery is sufficiently charged the green LED will light. The charger 

can be left connected for extended periods if required.
The charger also has a discharge facility built in which is triggered by the 

yellow button at the front of the charger. Use of this is not essential but it can 

be used to ensure a full discharge/charge cycle if required.

Our sampling pumps are designed to withstand the punishment of a long 

and active working life. They are manufactured from 2.5mm thick aluminium, 

cold formed and welded along all seams. This results is a case that is rigid, 

non rusting and durable. The units are finished in a tough epoxy powder 

coating which is resistant to minor damage.

Internally the pumps contain a high efficiency, rotary vane vacuum unit 

with a capacity of up to 3.0 litres/min of air through a standard 0.8 micron 

membrane filter.

The flow is set from 1.0 litres/min to 3.0 litres/min by means of an electronic 

control unit. This can be adjusted by turning the flow control screw found on 

the underside of the unit (6)

Flow control is maintained to or - 3% of the set figure by an 

electronic control unit which is epoxy sealed to protect against moisture, 

heat and vibration.

Power is provided by an 8 cell 9.6v 2000mah battery pack, this allows for 

approximately 8 hours sampling at 3.0 litres/min.

1) On/Off switch
2) Timer 
3) Inlet
4) Time Reset
5) Red & Green LEDs
6) Flow Control
7) Charging Socket

1

6

3

5

    00

4

2

7

Pump will not start, red LED 

illuminated.

Battery voltage too low, recharge 

battery.

Flow rate insufficient.

Check motor speed control 

setting and sample tube for 

kinks.

Pump stops while running, red LED

illuminated.

Battery voltage has dropped to 

5.5v, rechargeor replace battery.

Pump fails to run for specified duration. Check battery pack.

Troubleshooting