SOLR
®
100% OXYGEN MASK LEVEL 3
TM0063-2
60410X
5-11
27.
Using compressed air, aim the nozzle at the set screw opening and blast with compressed air to remove
the balancing chamber.
Figure 5-30 Remove Balancing Chamber
28.
Using a plastic pick, remove the two O-rings ID .228 from the balancing chamber and discard.
29.
Place balancing chamber and the regulator body to the side.