OPERATION

To operate 

A/T SCANX ERASER.

1. Open the 

A/T SCANX ERASER 

lid.

2. Place Phosphor Storage Plates (PSP’s) on 

A/T SCANX ERASER

tray with blue or white side of PSP’s 

facing up as suggested in FIG. 3 - 8.

3. Close the 

A/T SCANX ERASER 

lid.

4. The 

A/T SCANX ERASER

will automatically start the erasing process and run for approximately 35 

seconds to completely erase the PSP’s. The Illumination Indicator will be lit during the erasing process 
and will turn off upon its completion.

Cleaning Instructions:

1. Disconnect the A/T SCANX ERASER by removing the supplied line cord from the unit.
2. Wipe the glass surface underneath the lid, and the grooves of the tray using a clean, damp cloth.
3. Wipe the surfaces again until they are dry with a dry cloth.
4. Once the cleaned surfaces have dried, reconnect the supplied line cord to the A/T SCANX ERASER.

PHOSPHOR STORAGE PLATE (PSP) HANDLING

Important: Completely erase Phosphor Storage Plates (PSP’s) prior to use.

Always allow the PSP’s to remain in Eraser until completion of cycle.

Use PSP’s within 24 hours of last ersasure. Repeat erasing process if PSP’s have been stored longer 
than 24 hours.

Handle PSP’s with care.

- Avoid scratching or soiling PSP’s.
- Do not bend PSP’s or apply unnecessary pressure.
- Do not store PSP’s in a hot or moist area.
- Protect the PSP’s from direct sunlight and ultraviolet rays.
- Pick up the PSP’s using two fingers around the edges to avoid unnecessary contact with the plates.

Disposal of PSP’s.

Consult with your federal, national, state and local government, for rules and regulations on
disposal of PSP’s.

Summary of Contents for SCANX A

Page 1: ...A T SCANX ERASER OPERATOR S MANUAL...

Page 2: ...CONTENTS Introduction 3 Warranty 3 Safety Instructions 3 Specifications 4 Dimensions 5 Installation 5 Operation 6 Cleaning Instructions 6 Phosphor Storage Plate Handling 6 Phosphor Storage Plate Place...

Page 3: ...nties expressed or implied No representative or person is authorized to assume for us any liability in connection with the sale of our equipment WARRANTY 3 DO NOT ATTEMPT INTERNAL SERVICE The interior...

Page 4: ...Pollution Degree 2 Erasing Surface Size Erasing area 8 x 12 Phosphor Storage Plate Capacity All sizes up to 21 Size 2 PSP s See PHOSPHOR STORAGE PLATE PLACEMENT for examples of more options Light Out...

Page 5: ...e guidelines for optimal performance Place on a stable flat counter top large enough to hold the Eraser and to provide a working area Provide access to a standard grounded electric 115V AC outlet with...

Page 6: ...Wipe the surfaces again until they are dry with a dry cloth 4 Once the cleaned surfaces have dried reconnect the supplied line cord to the A T SCANX ERASER PHOSPHOR STORAGE PLATE PSP HANDLING Importa...

Page 7: ...EMENT FIG 4 ARRANGEMENT FOR 21 SIZE 2 PSP s FIG 8 ONE 8 x 10 CEPH PSP FIG 5 ONE 5 x 12 PAN PSP FIG 7 ONE 15cm x 30cm PAN PSP FIG 6 ONE 5 x 12 PAN PSP FOUR SIZE 2 PSP s FIG 3 SUGGESTED PSP ORIENTATION...

Page 8: ...T VistaCam Omni T VistaCam Omni ic4 Air Techniques is a leading manufacturer of fine dental equipment from air and vacuum systems and x ray film processors to an impressive line of new products incor...

Reviews: