background image

Alarms 

Alarm Types: 

Screen Display 

Detail 

LOW ALARM 

Audible Alarm: One (1) slow 
beep every second 
Visual Alarm: One (1) slow 
flash every second 
Vibrating Alarm: One (1) slow 
vibration every second 

HIGH ALARM and OVER 
LIMIT (OL) ALARM 

Audible Alarm: Two (2) fast 
beeps every second 
Visual Alarm: Two (2) fast 
flashes every second 
Vibrating Alarm: Two (2) fast 
vibrations every second 

DETECTOR LIFE 
COUNTDOWN ALARM “EOL” 

Once the unit has less than 
One (1) month of life remaining, 
the unit will switch to days 
remaining, when it is less than 
One (1) day remaining it will 
switch to hours remaining. 
Once the detector has Eight (8) 
hours remaining, it will begin to 
beep, flash, and vibrate 
intermittently. To end the 
alarms press the button down. 
Once the detector has reached 
the end of its operating life the 
display will show “EOL” (End Of 
Life). 

Alarm Set Points: 

Default factory set points: 
H2S: Low Alarm 10 ppm/High Alarm 15 ppm  
CO: Low Alarm 35 ppm/High Alarm 200 ppm 
O2: Low Alarm 19.5%/High Alarm 23.5% 

*Note these set points can be changed using the GCT IR
Link. Please refer to the SGC IR Link documentation for 
further details. To display the detector alarm set points press 
the button on the front of the detector. 

 

Use caution when changing alarm set points. 
Confirm these levels with your company safety 
officer. 

 

DO NOT use IR communications when an 
explosive atmosphere may be present. 

Event Log 

By default the detector stores the last twenty five (25) alarm 
events. The system stores events by first in first out, i.e. the 
26

th

 event will replace the first event and so on. This 

information can be downloaded using the GCT IR Link. For 
each alarm event the detector records the following: 

The detector serial number

Bump Test (Yes or No)

Life remaining on the detector

Number of self-tests performed

Number of events

Alarm Condition (Low, High, or OL)

Specific event date and time

Peak gas concentration in ppm or %.

Bump Test Interval (bUP) 

Using the IR Link or GCT Manager, units can be 
programmed to alert the user if a bump test is due. This 
interval can be set anywhere from 1 to 365 days. 

*Note the 

unit default is to have no bump interval programmed.

 

If a detector is due for a bump test, the unit display will 
alternate between the months remaining and “bUP”. In 
addition the unit will emit alternating flashes (left and right) 
every 5 seconds. And the “test” icon will remain even after a 
button push. 

This alert can be cleared by either placing the unit in a Clip 
Dock or, if the bump interval is set, by manually applying gas 
to the unit. To manually clear the alert, press the button 
down once and wait for the "GAS" to show on the display 
while the TEST icon flashes. The detector will wait for 45 
seconds for the target gas to be applied, or

 

a button press to 

skip the bump test.

 

If the SGC is bumped while showing 

"GAS", then it will record as a bump test in the event log 
instead of as an exposure. If no gas is applied, it will return 
to the normal screen and will not record anything in the 
event log.

Single Gas Clip and SGC Plus User’s Manual Rev 

2.05

Single Gas Clip and SGC Plus User’s Manual Rev 

2.05

Summary of Contents for Single Gas Clip

Page 1: ...ns H2S 25 ppm CO 200 ppm O2 18 Bump Test can be performed either manually or through the Clip Dock If a manual test is to be performed make sure to test in a clean air environment If a unit fails the Self test or Bump Test please discontinue use of the detector The detector contains a lithium battery that must be disposed of by a qualified recycler Do not substitute internal components as this may...

Page 2: ...ns with a maximum character limit on the User ID of 6 characters The User ID can be changed modified via the IR Link software Screen 5 if applicable If the detector has been exposed to gas exceeding the low alarm set point a value will appear with max next to it This represents the peak value highest that the detector has seen After this screen there will be another screen displaying a value with ...

Page 3: ...events The system stores events by first in first out i e the 26th event will replace the first event and so on This information can be downloaded using the GCT IR Link For each alarm event the detector records the following The detector serial number Bump Test Yes or No Life remaining on the detector Number of self tests performed Number of events Alarm Condition Low High or OL Specific event dat...

Page 4: ...n be turned off Hibernated to suspend the 24 month operation life countdown SGC Plus Hibernation with the IR Link 1 Check that you have installed the IR Link Software and the IR Link USB connections are plugged in 2 Click Read Device on the IR Link Software 3 Note when the monitor is hibernated the event log will be cleared It is highly recommended to save the event log by pressing the save event ...

Page 5: ...nt shall GCT s liability hereunder exceed the purchase price actually paid by the buyer for the product This Warranty does not include 1 Fuses disposable batteries or routine replacement of parts due to the normal wear and tear of the product arising from use 2 Any product which in GCT s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation handling ...

Reviews: