ISO 14117:2019
Clause 4.9 - Test
requirements for
the frequency
range of 385 MHz
≤ ƒ ≤ 3000 MHz
Per standard
Functions as it did
before the test
without further
adjustment after
application of the
test signal per
standard
Seek the advice of your
physician or other qualified
health provider regarding
Transmitting Devices
and
Cellular and Mobile
Phones
•
Exercise caution in
the vicinity of
equipment that
generates strong
radio-frequency fields.
•
Do not enter an area
with posted warnings
advising pacemaker
patients (or patients
with other types of
implantable devices)
not to approach.
•
Interference may
occur in the vicinity of
equipment marked
with the following
symbol:
ISO 14117:2019
Clause 5 -
Testing above
frequency of 3000
MHz
Standard does
not require testing
of devices above
3 GHz.
Electromagnetic
fields > 3 GHz are
not expected to
interfere with
device operation
because of the
increased device
protection
afforded by the
attenuation of the
enclosure and
body tissue at
microwave
frequencies, the
expected
performance of
EMI control
features
implemented to
meet lower-
N/A
Avoid direct exposure to the
main lobe of high-power
radar and microwave
communication beams.