Relays 

6.2 Trigger inputs

The 

trigger inputs 

TR8 and TR9 can be driven either via optocouplers by volt-

age levels (4 to 30 V) or via the two keys 

F1

 and 

F2

 (configuration, see 6.3.).

When using floating switch contacts the optocouplers must be appropriately
wired with supply U+ and U- 
(see diagram). 
The trigger function (as standard to start
or stop a measuring operation) can also
be freely configured (see Manual 6.10.9).

6.3 Analog outputs

The relay adapter can also - in various options - be fitted with electrically iso-
lated 

analog outputs 

offering the following signals. 

Option            

Output signal 

Gain 

OA 8000-R2

0.000 to +10.000 V

0.5 mV / digit

or

0.000 to +20.000 mA

1   µA / digit

The output value normally corresponds to the measured value for the selected
measuring point. Or alternatively the analog value can be specified as control
variable via the interface (see  Manual 6.10.7). The output signal can be pro-
grammed as standard output 0 to 10 V, 0 to 20 mA, 4 to 20 mA for any partial
measuring ranges (see Manual 6.10.7).

6.4 Connecting peripheral equipment

Peripherals can be connected via the supplied ALMEMO screw connector  ac-
cording to the following schematic diagram :

 ALMEMO

®

 8006-RTA3

5  

Summary of Contents for ALMEMO 8006-RTA3

Page 1: ...____________________________ Operating instructions Relay trigger adapter analog ALMEMO 8006 RTA3 V2 4 18 03 2014 www ahlborn com...

Page 2: ...uts TR8 TR9 2 DC socket Mains adapter ZA1312 NA1 12V 0 2A 3 LCD graphics display 7 rows for functions 1 line for softkeys F1 F2 Shown in brackets MENU 4 Operating keys TR8 9 Trigger keys softkeys P Po...

Page 3: ...4 Connecting peripheral equipment 5 6 COMMISSIONING 6 7 DISPLAY AND OPERATING CONTROLS 7 7 1 Main menu 7 7 2 Showing all elements 7 7 3 Showing single elements 7 7 4 Device configuration 8 7 5 Messag...

Page 4: ...maximum supply current on the measuring instrument must take the sensors into consideration If the maximum supply current is exceeded a mains adapter e g ZA1312 NA1 should be connected at the DC socke...

Page 5: ...with electrically iso lated analog outputs offering the following signals Option Output signal Gain OA 8000 R2 0 000 to 10 000 V 0 5 mV digit or 0 000 to 20 000 mA 1 A digit The output value normally...

Page 6: ...uring points program limit values on the ALMEMO device 2 With ALMEMO device program the first relay port to variant 0 Summated alarm using command i20 f9 k0 see Manual 6 10 9 2 3 In the event of any l...

Page 7: ...appro priate measuring point see Manual 6 10 3 8 Display and operating controls The adapter incorporates a graphics display that can show exactly all states affecting the interface elements This is c...

Page 8: ...desired function can be selected by means of keys PROG and Programming is by means of the softkeys ON OFF or SET and input is by means of keys PROG PROG Illumination To switch display illumination ON...

Page 9: ...Watchdog Duration s o Duration 060s Switch off watchdog alarm with watchdog function and switch on again after restoring the control 8 5 Messages Each relay can be assigned a particular message either...

Page 10: ...output port pp ipp f9 ayyyyy Program message for port pp e g ipp message port pp Start new line with special character Tel 123 CRLF To output message Response ipp P48 Message port pp Tel 123 CRLF To...

Page 11: ...ty 50 V 0 5 A Trigger inputs Optocoupler 4 to 30 V Input current 2 mA Analog outputs Electrically isolated switchable OA 8006 R02 4 00 to 10 0 V 0 5 mV digit Load 100 kW 0 0 to 20 0 mA 1 A digit Load...

Page 12: ...mp connectors ZA 8006 RTA3 Mains adapter with ALMEMO connector 12 V 1 A ZA 1312 NA8 Options 2 additional relays maximum 10 including 1 ALMEMO clamp connector OA 8006 SH2 2 additional normally closed r...

Page 13: ...nection diagram 2 2 contact assignment 8 3 8 Current consumption 11 1 11 Device configuration 8 4 8 Display and operating controls 8 7 display illumination 8 4 8 driven inversely 9 10 Driving the anal...

Page 14: ...Power supply 5 4 Power supply 11 1 11 Product overview 11 2 12 Programming via the device interface 9 10 Relay 11 1 11 relays 8 3 7 screw connector 6 4 5 Showing all elements 8 2 7 Showing single elem...

Page 15: ...Index trigger source 8 3 8 variant 8 3 7 voltage capacity 6 1 4 Watchdog 8 4 9 8 4 8 ALMEMO 8006 RTA3 15...

Page 16: ...11 Appendix 11 4 your contact 16 ALMEMO 8006 RTA3...

Reviews: