background image

  5. Introduction 

Measuring ranges 

Appropriate measuring ranges are available for all sensors with a non-linear
characteristic, e.g. 7 thermocouple types, NTC probes, and flow transducers
(rotating vanes, thermoanemometers). For humidity sensors additional function
channels are available for  calculating humidity variables such as dew point,
mixture ratio, vapor pressure, and enthalpy. Measured values from other sen-
sors can also be acquired using the voltage and current ranges with individual
scaling in the connector. Existing sensors can also be used - so long as the ap-
propriate ALMEMO

®

 connector is connected via its screw terminals. For digital

input signals, frequencies, and pulses, adapter connectors are available with
an integrated microcontroller. It is thus possible to connect numerous sensor
types to any ALMEMO

®

 measuring instrument and to change sensors without

the need for any extra settings. 

Function channels 

Maximum, minimum, and average values of certain measuring points can be
programmed as function channels and can be processed and printed like nor-
mal measuring points. 

Units 

The 2-character units display can be adapted for each measuring channel so
that both the display and the printout always indicate the correct  units,  e.g.
when a transmitter is connected. Conversion between °C (Centigrade) and °F
(Fahrenheit) is performed automatically. 

Measured value designation 

Each sensor is identified by means of a 10-character alphanumeric name. This
name is entered via the interface and will appear in the printout or on the com-
puter display. 

Correction of measured values 

The   measured  value   on  each  measuring  channel  can  be  corrected  both  in
terms of zero-point and gain; this means that even sensors usually requiring
initial adjustment (e.g. expansion, force) can be freely interchanged. Sensors
with   multi-point   calibration   can   also   be   connected;   (see   Manual   Section
6.3.13). 

Scaling 

The corrected measured value on each measuring channel can also be further
scaled in terms of zero-point and gain - using the base value and factor. The
decimal point position can be set by means of the exponent function. 

Limit values and alarm 

Per measuring channel two limit values can be set (1 maximum  and 1 mini-
mum). In the event of one of these limit values being exceeded relay output
modules actuate the associated alarm contacts; these can be allocated individ-
ually to specific limit values. Hysteresis is set by default to 10 digits but this can
be adjusted to any number between 0 and 99. The exceeding of a limit value
can also be used to start or stop measured value recording automatically. 

 10              ALMEMO

®

 2450-1

Summary of Contents for ALMEMO 2450-1

Page 1: ...____________________________ Operating instructions Universal measuring instrument ALMEMO 2450 1 V2 3 06 11 2013 www ahlborn com...

Page 2: ...V not type L Mains adapter ZA1312 NA1 12V 0 2A Cable el isol ZA2690 UK 10 30V Interface RS 485 option 5 LCD Function field a Function b Measuring point 2nd meas value c Units for 2nd measured value Ma...

Page 3: ...voltage monitoring 14 7 2 Mains operation 14 7 3 External DC voltage supply 14 7 4 Sensor supply 14 7 5 Switching ON OFF reinitialization 14 7 6 Data buffering 15 8 CONNECTING THE TRANSDUCERS 15 8 1...

Page 4: ...ddress and networking 24 12 2 Analog output 25 12 3 Switching OFF automatically 26 12 4 Device locking 26 13 TROUBLE SHOOTING 26 14 DECLARATION OF CONFORMITY 28 15 APPENDIX 28 15 1 Technical data 28 1...

Page 5: ...l ity tests We provide a guarantee lasting two years from delivery date that your device will function trouble free Before you send your device to us please observe the advisory notes in Chapter 13 Tr...

Page 6: ...l of any such item as unsorted do mestic waste is strictly forbidden Please dispose of all packaging materials according to the applicable national waste management regulations Please dispose of cardb...

Page 7: ...or conditions for which it is not suited Use of the device with an unsuitable power supply and or in conjunction with unsuitable peripheral equipment Use of the device for any purpose other than that...

Page 8: ...uitable Be sure to observe the maximum load capacity of the sensor power supply Sensors with their own integrated power supply are not electrically iso lated from one another 4 2 Handling batteries re...

Page 9: ...plete list of interface commands with all the printouts Manual Ch 7 The operating instructions you are now reading cover only those features and controls that are specific to this device Many sections...

Page 10: ...t always indicate the correct units e g when a transmitter is connected Conversion between C Centigrade and F Fahrenheit is performed automatically Measured value designation Each sensor is identified...

Page 11: ...most sensors sensor breakage is detected automatically except for con nectors with shunt dividers or additional electronics Analog output and scaling Each measuring point can be scaled by means of ana...

Page 12: ...g over measuring point scans The measured values from measuring point scans can be averaged either over the whole measuring duration or over the specified cycle These average val ues can then be outpu...

Page 13: ...ence value by pressing key CLR 6 Revert to normal measured value by pressing and holding down key CLR 7 Evaluating a measuring operation Call up maximum minimum values by pressing keys MAX MIN 6 To de...

Page 14: ...d with power from an external source preferably using mains adapter ZA 1312 NA1 12 V 0 2 A via the DC socket 4 Please ensure that the mains voltage is correct 7 3 External DC voltage supply The DC soc...

Page 15: ...isting sensors to ALMEMO instruments see Manual Ch 4 All suitable sensors with an ALMEMO connector see above have the measuring range and units al ready programmed and can thus be connected to the inp...

Page 16: ...can flow between sensors power supply and peripherals All points must therefore lie at the same potential and or any un equal potentials must be electrically isolated Sensors combined within one conne...

Page 17: ...digits b for depicting various measuring functions a there are also 2 arrows f for depicting the operating status Function field Main field Display of measuring functions in the function field Measur...

Page 18: ...points Measuring range function not permitted Multiplexer M5 D B is not permitted Sensor breakage flashes Outside of measuring range undershoots cold junction compensation flashes or cold junction com...

Page 19: ...Section 5 1 1 10 1 Measured value After switching ON first of all a segment test is performed then the battery voltage ap pears and if the batteries are almost empty 3 8 V the LOBAT arrow also appear...

Page 20: ...olt 2 ZA 9000 FS2D 260 00 260 00 mV d260 Difference volt ZA 9000 FS3D 0 2600 2 6000 V d2 60 Sensor voltage any 0 00 20 00 V UbAt Milliampere ZA 9601 FS1 0 000 26 000 mA I032 Percent 4 to 20mA ZA 9601...

Page 21: ...ature value can at the same time be dis played in the function field Select 2nd channel Activate temperature display Press and hold dow M Return to channel display Press and hold down M 10 2 Peak valu...

Page 22: ...this channel The MAX MIN and MEM functions are thus also available for relative measurement 11 OUTPUTS The following interfaces corresponding accessories or options either are nec essary see 14 2 11 1...

Page 23: ...e entered Or alternatively there are variants 2490 xR02 with 2 integrated electrically iso lated analog outputs see Section 15 2 these can be configured as 0 to 10 V 0 to 20 mA or 4 to 20 mA These app...

Page 24: ...ling for Analog output P0 7 option see 12 2 Automatic switch OFF time in minutes see 12 3 To enter a value first press ON and the value starts flashing To modify the value press keys M or M To delete...

Page 25: ...A for 30 0 to 120 0 C For the previously specified reference channel the values for analog output start and analog output end plus analog output type 0 to 20 mA or 4 to 20 mA can all be programmed Fir...

Page 26: ...o grammed in many versatile ways It is suitable for connecting a wide variety of different sensors additional measuring instruments alarm signaling devices and peripheral equipment Given these numerou...

Page 27: ...anual 6 2 1 Enter ctrl Q for XON if the device is in the XOFF status Check the programming by means of P15 see Manual 6 2 3 Test the transmit line by selecting a measuring point using com mand Mxx and...

Page 28: ...owing advisory notes must be observed when operating the device Using the device in strong electromagnetic fields may aggravate measuring er rors 50 V at 3 V m and 1 5 meters thermocouple sensor After...

Page 29: ...mm ABS acrylonitrile butadiene styrene weight approx 260 g Suitable conditions Operating temperature 10 to 50 C storage temperature 20 to 60 C Ambient relative humidity 10 to 90 rH non condensing 15...

Page 30: ...conductor relays trigger input ZA 1006 EAK 15 3 Index Accessories 15 2 30 additional channels 8 2 15 AMR Control 5 1 3 13 analog output 11 2 23 Analog output 12 2 25 analog output end 12 2 25 analog o...

Page 31: ...8 3 16 Power supply 15 1 14 29 Process control 5 1 3 12 Product overview 15 2 29 reference channel 12 2 25 reference value 10 4 22 reinitialization 7 5 14 Relative measuring 10 4 22 RS485 interface 11...

Page 32: ...3 6 WIN Control 5 1 3 13 15 4 Your contact AHLBORN Mess und Regelungstechnik GmbH Eichenfeldstra e 1 83607 Holzkirchen Germany internet http www ahlborn com e mail amr ahlborn com We take every possib...

Reviews: