INFRARED THERMOMETER
+49 2173 -164-0 | FREE FAX 0800 / 2 42 32 93 | [email protected]
12
13
WWW.AHB-SHOP.COM
SOL
UT
IONS
SOL
UT
IONS
INFRARED THERMOMETER
TEMPERATURE UNIT SETTING –
F1
SETTING OF TEMPERATURE ALARM POINT –
F2
TEMPERATURE OFFSET SETTING –
F3
PROMT FOR ON/OFF SETTINGS –
F4
TIP:
Hold down the “set” button for 2 seconds. The screen displays F 1 figure 8.3. Press the
“+” button again to select the Celsius or Fahrenheit temperature unit figure 8.4 figure
8.5.
Hold down the “set” button for 2 seconds. The screen displays F 1, press the “set” but-
ton to enter F 2 once, then press the “ten” button to increase the temperature by
0.1°C - 0.1°F, and the “–“ button to reduce the temperature by 0.1°C - 0.1°F. Note: The
default alarm value is 38°C - 100.4°F
Hold down the “settings” button for 2 seconds. The screen displays F 1, press the “set-
tings” button twice enter F 3, then press the “ten” button to increase the difference by
0.1°C - 0.1°F, press “–“ button to reduce the difference by 0.1°C - 0.1°F. If the weather or
environment changes, check and adjust the thermometer.
Hold down the “settings” button for 2 seconds. The screen displays F 1, press the “set-
tings” button 3 times to enter F 4, press the “ten” button to open the “OFF” symbol on
the buzzer screen or to close the buzzer symbol FF on the screen.
After entering the settings menu operation, if you need to modify a certain setting, you
can press the menu key to select continuously.
Save the new settings after the completion of the sequence settings, halfway shut-
down quit, the new settings will not be saved.
After enterning the setting menu, press the measure key is invalid, the thermometer
will not take the measure
1. Body temperature patterns are used to measure body temperature and are obtained
by dynamically compensating for the ambient temperature and the surface temper-
ature of the forehead.
(P 8.3)
(P 8.4)
(P 8.5)
HINT !
1. Before an after use, please keep the sensor and probe cavity clean.
2. Please place the thermometer in a temperature stable environment. When the
ambient temperature changes significantly, such as from indoor to outdoor, please
leave it for about 10 minutes to measure.
3. Do not take the temperature of an extremely hot or cold object immediately after
taking it. Leave it for 10 minutes before taking it.
4. If the test person comes from a place with a large temperature difference from the
measurement environment, the test person should remain in the test environment.
Body temperature
LOG
METHOD OF MEASUREMENT
Make sure the top of the screen is in temperature mode. Position the thermometer
probe in the middle of your forehead and keep it perpendicular, about 3 – 5 cm away.
Figure 9.1. Press the measure key, after about 1 second “beep”, display the measure val-
ue figure 9.2, the measure is finished. If the measured value exceeds the temperature
alarm point default value of 38°C, then issued “di: Di Di” five tone alert sound.
When ambient temperature changes significantly affect the forehead temperature,
please measure figure 9.3 behind the earlobe.
3 to 5 cm
3 to 5 cm
2. According to the environment temperature, measuring distance and individual skin
difference, the temperature offset value can be corrected by the difference between
the measured value and the true value. The range of corrections is –3.0°C ~ +3.0°C,
factory set to 0.0°C.
For example, if a thermometer measures a body temperature of 36.2°C and the tar-
get’s actual body temperature is 37.0°C, then it enters F 3, with a deviation of. Adjust
the temperature up to 0.8°C, an the result is the same as the actual temperature.
(P 9.1)
(P 9.2)
(P 9.3)