16
EN
RF emissions CISPR 11
Class B
The nasal aspirator
AGU NS15
is suit-
able for use in all establishments, includ-
ing domestic establishments and those
directly connected to the public low-volt-
age power supply network that supplies
buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
GUIDANCE AND MANUFACTURER’S
DECLARATION ELECTROMAGNETIC IMMUNITY
The
nasal aspirator
AGU NS15
is intended for use in the electromagnetic
environment specified below.
The customer or the user of the
nasal aspirator
AGU NS15
should assure that
it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV
contact
+ 8 kV air
Floors should be wood,
concrete or ceramic tile.
If
floors are covered with
synthetic material, the rela-
tive humidity should be at
least 30 %
Power frequency
(50/
60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char
-
acteristic of a typical location in
a typical commercial or hospi-