Installation-, Operating-, Maintenance-Manual, Spare part List
BCR05
AGT-PSG GmbH & Co.KG
Richard-Lucas-Str. 6, D-41812 Erkelenz, Deutschland
Tel: +49 (0)2431/9627-0, Fax: +49 (0)2431/9627-31
www.agt-psg.de, [email protected]
Technical Data
Gas paths
1
2
4
Heat exchanger
1 x MONO
2 x MONO
2 x DUAL
Material
PVDF
Glass
SS316
PVDF
Glass
SS316
PVDF
SS316
Gas flow
1)
L/h
250
300
400
2x250
2x300
2x350
4x125
4x150
max. inlet dew point
65
70
80
65
70
80
65
80
max. inlet temperature
°C
130°C
Outlet temperature
°C
3°C +/-0,5°C
Pmax with cond. pump
Follow manufacturers instructions
Pmax without cond. pump
bar
2,5
2,0
100,0
2,5
2,0
100,0
2,5
100,0
Gas connection
mm
Gas inlet/outlet DN4/6, condensate outlet D12
Dead space
ml
67
98
67
2x67
2x98
2x67
4x55
4x55
Ambient temperature
°C
+5°C to +45°C
Cooling power
W
300W at 25°C ambient temperature
Protection rate
IP20 acc. EN 60529 / EN 61010
Certifications
ATEX EX II 2G Ex px de [ai] IIC T4 Gb, for Zone 1 and 2
Housing
Console mounting / RAL 7035
Dimensions
mm
440 x 350 x 470 (WxHxD)
Temperature monitoring
Analog display and isolated alarm relay contacts
Power supply
230V 50/60Hz or 115V 50/60Hz
Power consumption
W
220W at 230VAC – starting current 6,3A
Weight
kg
38,5kg
40,5kg
1)
at standard conditions, dew point 65°C inlet temperature, 10-25°C ambient temperature