15
7.
After completing the X calibration, press the
>>l
key (11D) to
change the axis and to calibrate the Y axis. On the remote, press
the double arrow key
>>l
.
8.
If the Y axis doesn’t have to be calibrated, press the Auto/Man
key (6) to save the calibration you've just made on X axis. Or you
can save the calibration by pressing and holding for a few seconds
the rotation key (11A). On remote, use key with the small dot.
If you are not sure of the calibration, don’t save it and turn the
laser off using the On/Off key.
4.2.3 Checking Y axis
1.
Rotate the laser 90 degrees so that Y is facing the wall.
2.
Mark the location of the center of the beam.
3.
Rotate the laser 180 degrees so that Y’ faces the wall.
4.
Mark the location of the beam center near the first mark.
5.
At 100 ft., the marks should be no more than 3/16” apart (at
30m, no more than 5mm apart). This is within the stated accuracy of
± 3/32” at 100 ft. (± 0.0075%).
6.
If the marks are close enough, Y axis is within calibration.
Proceed to “Final X to Y Check.” If the marks are not close enough,
Y axis needs to be calibrated.
4.2.4 Calibrating Y axis
The laser must be calibrated to bring the beam to the center of the
two Y marks. Read "Calibration Overview" before proceeding.
If you are still in calibration mode from the X axis, turn Y’ towards
the wall and press the
>>l
key (11D) to change to the Y axis.
When the Y LED blinks slowly, it’s ready to be calibrated in Y axis.
If you’re no longer in calibration mode
1.
Turn off the laser.
2.
Hold the Auto/Man (6) key. While holding, momentarily press
the ON key.
3.
After the LEDs blink in sequence, release the Auto/Man key.
4.
Select the Y axis by pressing the
>>l
key (11D) on the laser
keypad. On the remote, press the double arrow key
>>l
.