Aginova Sentinel Next Series Reference Manual Download Page 2

                                                                                                                                                                                                      

Page 2 of 8 

 

 

2.

 

Power Button 

 

This is just a momentary push that functions as an On/Off button. 
To power the unit on – briefly press and release the power button.  The Mode LED (3) will blink 
green 5 times very quickly indicating the unit is turning on. 
 
To power the unit off- press and hold the power button for 5 seconds.  When you release the 
power button the Mode LED (3) will blink red 5 times very quickly indicating the unit is turning 
off. 

 

3.

 

Mode Button / LED 

 
The Mode Button / LED has several functions.   

LED indicator: 

Green (solid):  Sensor is in factory default Direct Mode 

Blue (periodic blink):  Sensor is configured to post data to a server. 

Blue (solid):  Sensor is configured to post data to a server and is plugged into a power 
source. 

5 quick green blinks:  Sensor is powering on. 

5 quick red blinks:  Sensor is powering off. 

Blinking red at same time as Battery Status LED (1):  Battery is almost 100% 
depleted.  Sensor cannot function until battery is recharged (rechargeable version) 
or replaced (battery only version).   

Red blinks per dbg table included discussed in Troubleshooting. 

 

Switching Modes: 

 

To return sensor to factory default Direct Mode, press and hold the Mode Button for ~5s.  

Upon release the Mode Button LED will flash green a few times and stay solid Green. 
 

To return sensor from Direct Mode to post to last configured data server, briefly press the 

Mode Button.  The Mode Button LED will flash blue a few times indicating it has been set to 
report to last configured data server. 
 

Control LCD Display (7) functions on units with display.

   

This function is covered in detail in section 7.  LCD Display Function. 

 
        

Control debug function. 

 

Covered in section 8.  Troubleshooting. 

 

Summary of Contents for Sentinel Next Series

Page 1: ...cable is plugged in to the Micro USB input 4 and connected to a charger the Battery Status LED will have the following behavior Solid Green Battery is fully charged Solid Red Battery is charging If bo...

Page 2: ...quick green blinks Sensor is powering on 5 quick red blinks Sensor is powering off Blinking red at same time as Battery Status LED 1 Battery is almost 100 depleted Sensor cannot function until battery...

Page 3: ...securely and completely Use only official Sentinel Next probes from Aginova Below is pictured an example of a probe that is properly connected 6 Internal Wi Fi Antenna The Sentinel Next communicates...

Page 4: ...o not know which version you have contact Aginova support with the sensor ID and they can inform you whether you have Version 1 or Version 2 The following instructions are applicable to Sentinel Next...

Page 5: ...the sensor Message On Mean sensor is switch On Off Sensor is switch Off OtaP Sensor is being re flashed over the air rst sensor was reset using the press button min and max values were reset ErrX disp...

Page 6: ...Unknown Error 6 Verify the following points Make sure that your WiFi Access Point AP can reach the internet Verify that the AP is correctly configured SSID channels DHCP etc You can use another device...

Page 7: ...des in fulfillment of its warranty obligation A replacement product or part including a user installable part that has been installed in accordance with instructions provided by Aginova assumes the re...

Page 8: ...Y PRODUCT UNDER THIS WARRANTY OR MAKE A PRODUCT EXCHANGE WITHOUT RISK TO OR LOSS OF PROGRAMS OR DATA OBTAINING WARRANTY SERVICE Please access and review the help resources provided in this documentati...

Reviews: