User’s and Service Guide U3047-90001
U3047AM12
Cal Kit Operational Check
Cal Kit Operational Check
The following procedure can be used to confirm the 1-Port Cal operation of an ECal module or Mechanical
Cal Kit with the U3047AM12 Test Set. The operation verification limits provided ensure that your
U3047AM12 and network analyzer are operating properly. Refer to
“Verifying Cal Kit Operational Check
to ensure the RF interface cables are installed and working correctly before returning
the test set for repairs.
Verification Limits
Specifications for the test set are typical. System performance for the network analyzer and test set are
only characteristic and intended as non-warranted information. Only a functional certificate is provided for
the test set.
It is recommended that you return your instrument to Agilent Technologies for servicing or repair if the test
set and network analyzer performance exceed the operational verification limits.
A periodic calibration is not required. The Operators Check should be performed after System Setup, or if
performance is in question. An N-Port calibration should be performed before making a measurement.
Refer to
NOTE
If you suspect that your 16-Port configuration is not operating properly, ensure that all front
RF jumper interconnect cables are correctly attached.
Table 12
System Verification Limits
1
1.
Reflection Tracking takes into account Source Loss, Receiver Loss,
Margin, and network analyzer Mixer Cal.
Frequency
Ports 1 to 4
Ports 5 to 12
Ports 13 to 16
10 MHz to 500 MHz
−
2
−
2
−
2
500 MHz to 3.2 GHz
−
3
−
3
−
3
3.2 GHz to 10 GHz
−
5
−
5
−
4
10 GHz to 16 GHz
−
5
−
5
−
4
16 GHz to 20 GHz
−
6
−
6
−
5
20 GHz to 26.5 GHz
−
8
−
7
−
6
26.5 GHz to 43.5 GHz
−
10
−
9
−
7
43.5 GHz to 50 GHz
−
14
−
12
−
8
50 GHz to 67 GHz
−
20
−
18
−
12