![Agilent Technologies U2000 Series Operating And Service Manual Download Page 100](http://html1.mh-extra.com/html/agilent-technologies/u2000-series/u2000-series_operating-and-service-manual_2867112100.webp)
82
Agilent U2000 Operating and Service Guide
Appendix
Appendix A: Zero Set, Zero Drift, and Measurement Noise
The following specifications in
to the U2000 Series USB power sensors with the serial
prefix as shown below:
Table 5-1
Zero Set, Zero Drift, and Measurement Noise for Average only
Mode
Model
Serial Prefix
U2000A
Serial Prefix MY470/SG470 and below
U2001A
Serial Prefix MY471/SG471 and below
U2002A
Serial Prefix MY472/SG472 and below
U2004A
Serial Prefix MY474/SG474 and below
Range
1
1 Condition: (i) 0 °C to 55 °C and (ii) 40 °C, 95% relative humidity.
Zero Set
Zero Drift
2
2 Within one hour after zero set, at a constant temperature, after 24-hour warm-up
of the power sensor.
Measurement
Noise
3
3 The number of averages at 16 for
Normal
speed, measured over one minute
interval and two standard deviations.
–60 dBm to –35 dBm
±651 pW
996 pW
1.91 nW
–38 dBm to –15 dBm
±1.13 nW
400 pW
2.24 nW
–20 dBm to –9 dBm
±12.8 nW
6.01 nW
40.8 nW
–11 dBm to –5 dBm
±445 nW
155 nW
1.63 µW
–7 dBm to 15 dBm
±4.26 µW
3.20 µW
861 nW
10 dBm to 20 dBm
±6.84 µW
3.39 µW
19.5 µW
Summary of Contents for U2000 Series
Page 1: ...Agilent Technologies Agilent U2000 Series USB Power Sensors Operating and Service Guide...
Page 14: ...xiv U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 16: ...xvi U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 18: ...xviii U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...