U1731A/U1732A Quick Start Guide
14
Capacitance (Parallel Mode), Test Frequency = 10 kHz
Inductance (Series Mode), Test Frequency = 100 Hz/120 Hz
Range
Maximum
Display
Accuracy
Specified Note
Capacitance
DF
50 µF
50.0 µF
3.0% + 8 (DF<0.1)
12.0% + 100/Cx + 10 (DF<0.1)
After short cal.
20 µF
19.999 µF
3.0% + 6 (DF<0.2)
5.0% + 100/Cx + 8 (DF<0.2)
After short cal.
2000 nF
1999.9 nF
1.5% + 5 (DF<0.5)
1.5% + 100/Cx + 6 (DF<0.5)
-
200 nF
199.99 nF
1.5% + 5 (DF<0.5)
1.5% + 100/Cx + 6 (DF<0.5)
-
20 nF
19.999 nF
1.5% + 5 (DF<0.5)
1.5% + 100/Cx + 6 (DF<0.5)
-
2000 pF
1999.9 pF
2.0% + 6 (DF<0.5)
3.0% + 100/Cx + 6 (DF<0.1)
After open cal.
200 pF
199.99 pF
3.0% + 8 (DF<0.1)
5.0% + 100/Cx + 8 (DF<0.1)
After open cal.
N O T E
1
Q value is the reciprocal of DF.
2
This specification is based on the measurement performed at the test socket.
3
DUT and test leads need to be properly shielded to GUARD if necessary.
4
Cx = Counts of displayed C value, e.g. C = 88.88 µF then Cx = 8888.
N O T E
It is recommended to set the test frequency to 1 kHz for multi layer ceramic capacitor 10 ìF
and below.
Range
Maximum
Display
Accuracy
Specified Note
Inductance
DF
1000 H
999.9 H
1.0% + (Lx/10000)% + 5
2.0% + 100/Lx + 5
After open cal.
200 H
199.99 H
0.7% + (Lx/10000)% + 5
1.2% + 100/Lx + 5
-
20 H
19.999 H
0.7% + (Lx/10000)% + 5
1.2% + 100/Lx + 5
-
2000 mH
1999.9 mH
0.7% + (Lx/10000)% + 5
1.2% + 100/Lx + 5
-
200 mH
199.99 mH
1.0% + (Lx/10000)% + 5
3.0% + 100/Lx + 5
After short cal.
20 mH
19.999 mH
2.0% + (Lx/10000)% + 5
10.0% + 100/Lx + 5
After short cal.