7
Performance Tests
158
Agilent U1602A/U1604A User’s and Service Guide
List of Test Equipment
For scope performance verification, the recommended test equipments are
listed as below.
Table 7-1
List of test equipments for scope performance test
Application
Equipment
Critical Specifications
Recommended Equipment
Voltage measurement
accuracy
Calibrator / Power supply
5 mV to 35 VDC, 0.1 mV resolution
FLUKE 5520A
Digital multimeter
Better than 0.01% accuracy
Agilent 34401A
Cable
BNC, quantity of 2
Agilent 10503A
Shorting capacitor
BNC
Agilent 1250-0774
Adapter
BNC (f) to banana (m)
Agilent 1251-2277
Adapter
BNC tee (m) (f) (f)
Agilent 1250-0781
Bandwidth
Calibrator
NA
FLUKE 5520A
Feedthrough
50&, BNC connector (m) and (f)
Agilent 11048C
Cable
Type N (m), 24 inches
Agilent 11500B
Horizontal time base
Calibrator
5 ppm after 1/2 an hour
FLUKE 5520A
accuracy
Feedthrough
50&, BNC connector (m) and (f)
Agilent 10503A
Cable
BNC, 3 feet length
Agilent 11048C
Trigger sensitivity
Calibrator
25 MHz, 60 MHz and 100 MHz sine
waves.
FLUKE 5520A
Power splitter
Outputs differ < 0.15 dB
Agilent 11667B
Cable
BNC, quantity of 3
Agilent 10503A
Adapter
N (m) to BNC (f), quantity of 3
Agilent 1250-0780
Feedthrough
50&, BNC connector (m) and (f),
quantity of 2
Agilent 11048C