Technical Information
SETPOINT Submenu
17
This submenu contains the settings related to the set point contact
output with relay.
TYPE
: the output signal can be setted referring on one of following
values:
-
FREQUENCY, default value
-
POWER drawn by the pump
-
TIME
-
NORMAL, when the pump has reaches Normal Operation mode
(Both HiFlow or HiComp)
SET VALUE
, after selection of the analog output type, select the
desired value from 0000 to 9999
HYSTERESYS
, if a Hysteresys value is required, it can be adjusted
between 00 and 99
MASK TIME
Selectable between 00:00:00 and 99:99:99
OUTPUT LOGIC
The contact logic is configured depending on
Activation Type parameter status.
-
If Activation Type is “high level active” the contact is configured
as “normally open” (it will close the circuit when activated);
-
if Activation Type is “low level active” the contact is configured
as or “normally closed” (it will open the circuit when activated).
TwisTorr Medium-TMP Rack Controller User Manual /
87-901-049-01 (B)
261/284