TriScroll 620 Series Vacuum Pump
DRAFT 10/30/14
Electrical Connections
Wire the motor and electrical interlocks (if applicable) in accordance with local electrical codes and the relevant electrical
component manufacturer’s instructions. Table 2 and Table 3 list the full load motor currents at various voltages.
* Only available on pump model X3862-64000.
Single Phase Motor Connection
The pump can be configured for low voltage or for high voltage. As supplied from the factory, the pump is configured for
low voltage. Figure 4 shows the electrical connections for a single phase motor.
Figure 4 Single Phase Motor Electrical Connections
1. Verify the electrical supply voltage.
2. Remove the three screws (item 12 in Figure 3 on
page 11) that are holding the motor electrical cover
(item 13 in Figure 3).
3. Refer to Figure 4 to connect the motor to match your
supply voltage.
Two options are available to strain relieve the electrical
supply cable.
❑
The cable can be mechanically held under the
electrical cover in the groove provided. The
groove is sized for 14 gage cable.
❑
A 1/2
14 NPSM hole is also provided next to the
motor cover.
4. Wire L1 and L2 per Figure 4 using right angle flag
connectors or ring connectors.
Ensure that no exposed wiring is close to the electri-
cal cover or to other terminals on the board.
5. Secure the ground wire under the ground screw using
a ring connector.
6. Replace the motor electrical cover and secure it with
the three (3) screws removed in step 2.
Table 2 Full Load Motor Currents, Amps
1 phase motor
100 V
115 V
200 V
230 V
50 Hz
14
13.6
7.0
6.8
60 Hz
15.2
12.6
7.6
6.3
Table 3 Full Load Motor Currents, Amps, Low
Voltage*
1 phase motor
94 V
104 V
188 V
208 V
50 Hz
14.6
14.0
7.3
7.0
60 Hz
16.8
14.4
8.4
7.2
5
.
4
.
1
.
2
.
High Voltage
.
.
Low Voltage
5
.
4
.
1
.
2
.
L1 Power Cord
L1 Power Cord
Blue (Internal)
VPI Valve (Optional)
L2 Power Cord
VPI Valve (Optional
)
Violet
(Internal)
Violet (Internal)
VPI Valve (Optional)
Blue (Internal)
L2 Power Cord
VPI Valve (Optional)
Summary of Contents for TriScroll 620 Series
Page 4: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 6: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 20: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank DRAFT 10 30 14...
Page 24: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 25: ......