General Information

Agilent TriScroll 600 series pumps are designed to provide years of trouble-free service if maintenance procedures and
intervals  are  observed.  Cleaning,  bearing  grease  replenishment  and  tip  seal  replacement  is  recommended  when  pump
base pressure has risen to an unacceptably high level for your application. Bearings, rotary seals and o-rings should also
be replaced if the pump exhibits humming or grinding noises form the bearings. Main bearing life may be shortened if
your application requires the pumping of high quantities of water vapor. Use of the bearing purge described earlier will
keep this water from impacting bearing life.

Maintenance should be performed in accordance with procedures, tooling and materials specified in the manuals listed
below.

Related TriScroll Manuals

Other manuals related to the Installation and Operation, Tip Seal and Pump Module Replacement for TriScroll 600 series
pumps are listed in Table 5.

Maintenance and Tooling Kits

Material and tooling required to perform maintenance on TriScroll pumps is provided in kit form. A description of each kit
and ordering information is provided in Table 6.

Table 5

Related TriScroll Manuals

Part Number

Title

Applicable TriScroll Model

699904290

Installation and Operation Manual

All TriScroll 600 models

699904295

Installation and Operation Manual

All TriScroll 610 models

699904310

Tip Seal Replacement Manual

All TriScroll 600 and 610 models

699904300

Pump Module Replacement Manual

All TriScroll 600 and 610 models

699904305

Maintenance Manual

All TriScroll 600 and 610 models

699904300

Maintenance Manual

All TriScroll 600 and 610 models

Table 6

Maintenance and Troubleshooting Kits

Part Number

Description

Contents

Applicable TriScroll 

Model

PTSS0600MK

Maintenance Kit

All bearings, bearing 
seals, bearing 
lubricant, O-rings, and 
tip seals required to 
rebuild TriScroll 600 
series pumps.

All TriScroll 600 and

610 models

PTSS0600TK

Maintenance Tool Kit

All fixtures and tools 
required to perform 
any maintenance on 
TriScroll 300 series and 
TriScroll 600 series 
pumps.

All TriScroll 300, 310, 
600 and

610 models

Summary of Contents for TriScroll 610

Page 1: ...D R A F T 4 1 4 1 1 INSTALLATION AND OPERATION MANUAL Vacuum Products Division TriScrollTM 610 Dry Scroll Vacuum Pump Manual No 699904295 Revision F February 2005...

Page 2: ...gilent Inc TriScroll 610 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Loctite and PST are registered trademarks of Loctite Corporation Krytox Teflon and Viton are registered trademarks o...

Page 3: ...ntents Contacting Agilent vi Specifications 1 Unpacking and Inspection 4 Installation 5 Safety 5 Electrical Connections 6 Operation 11 Start Up Procedure 11 Shutdown Procedure 11 Troubleshooting 12 Tr...

Page 4: ...TriScroll 610 Dry Scroll Vacuum Pump This page intentionally left blank D R A F T 4 1 4 1 1...

Page 5: ...ue se refiere esta declaraci n es conforme a la s norma s u otro s documento s normativo s waamaar deze verklaring verwijst aan de volende norm en of richtlijn en beantwoodt a cui se rifersce questa d...

Page 6: ...Agilent In the United States you can contact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web site at http www...

Page 7: ...nterface dimensions See Figure 2 on page 3 Peak pumping speed 50 Hz 420 I m 25 2 m3 hr 14 8 cfm 60 Hz 500 I m 30 m3 hr 17 6 cfm Ultimate pressure Torr 7 0 x 10 3 Torr 9 3 x 10 3 mbar Maximum inlet pre...

Page 8: ...1 00 HP 0 75 kW 3 phase 50 Hz 200 230 380 415 VAC 3 Phase 60 Hz 200 230 460 VAC Motor full load currents See Table 2 Motor thermal protection Type N automatic Operating speed 1725 RPM 60 Hz 1425 RPM...

Page 9: ...3 TriScroll 610 Dry Scroll Vacuum Pump Figure 2 TriScroll 610 Vacuum Pump Interface Drawing...

Page 10: ...removing the foam packing and this manual slit open the inner box Lift the pump with plywood base out of the inner box Remove the four bolts securing the pump frame to the plywood base Locate the NW2...

Page 11: ...se electric shock short circuits and severe bodily injury WARNING Before inspecting or servicing the pump be sure the electrical supply is disconnected Consult a qualified electrician when ever changi...

Page 12: ...e relevant electrical component manufacturer s instructions Table 2 lists the full load motor currents in Amps at various voltages Table 2 TriScroll 610 Full Load Motor Currents VAC 100 115 200 230 1...

Page 13: ...al cover in the groove provided The groove is sized for a 14 gauge cable A 1 2 14 NPSM hole is also provided next to the motor cover The clamp fitting is not provided 5 Wire L 1 and L2 per Figure 4 us...

Page 14: ...rotation With the intake uncovered a large gas flow should come out the pump exhaust If there is no flow out the exhaust reverse any two of the leads L 1 L2 or L3 Repeat the test to verify correct rot...

Page 15: ...Installation Data TriScroll Vacuum Pump VPI Valve Part NW25 VPI Valve Part NW40 Solenoid Operating Volt age Range Wiring Diagram 100 120 V 50 60 Hz 1 Phase VPI251205060 VPI401205060 90 132 V 50 60 Hz...

Page 16: ...clamp and inlet O ring 3 Remove three 3 M5 screws item 1 on Figure 1 on page 1 that secure the cowling to the scroll module Remove the cowling 4 Tilt the pump back so that the rear end of the motor to...

Page 17: ...g purge port item 6 on Figure 1 on page 1 is provided to protect the main crankshaft bearings in applications where large amounts of water are being pumped In the standard configuration the port is se...

Page 18: ...ectrical cover Inspect and repair Wiring loose or cut Repair or replace Excessive voltage drop Check size and length of cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak...

Page 19: ...0 series pumps are listed in Table 5 Maintenance and Tooling Kits Material and tooling required to perform maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering...

Page 20: ...Scroll Model Table 7 Factory Service Options Model Factory Service Options EXPPTS06001 Advance Exchange TriScroll 600 Single Phase Pump EXPPTS06003 Advance Exchange TriScroll 600 Three Phase Pump EXPP...

Page 21: ...the core product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product...

Page 22: ...ed or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled...

Page 23: ...Other Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Prod...

Page 24: ...TriScroll 610 Dry Scroll Vacuum Pump This page intentionally left blank...

Page 25: ......

Page 26: ...y Agilent Technologies Vacuum Products Division via F lli Varian 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Toll Free 00 800 234 234 00 Fax 39 011 997 9350 Japan Agilent Technologies Vacuum Produ...

Reviews: