Technical Information
Agilent TPS-compact User Manual
223
Window Meanings
N.
Read/
Write
Data
Type
Description
Admitted Values
000
R/W
L
Start/Stop (in remote/ mode the window is a read
only)
Start = 1
Stop = 0
001
R/W
L
Low Speed Activation (in remote/ mode the window is
a read only)
No = 0
Yes = 1
(default = 0)
008
R/W
L
Remote or Serial configuration
Remote = 1
Serial = 0
(default = 1)
100
R/W
L
Soft Start (write only in Stop condition)
YES = 1
NO = 0
101
R/W
N
R1 Set Point type
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 0)
102
R/W
N
R1 Set Point valve (expressed in Hz, W or s)
(default = 867)
103
R/W
N
Set Point delay: time between the pump start and the
set point check (seconds)
0 to 999999
(default = 0)
104
R/W
L
Set Point signal activation type: the signal can be "high
level active" or "low level active"
0 = high level active
1 = low level active
(default = 0)
105
R/W
N
Set point hysteresis (in % of value)
0 to 100
(default = 2)
107
R/W
L
Active Stop (write only in stop)
0 = NO
1 = YES
(default = 0)
108
R/W N
Baud
rate
600
=
0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
111
R/W
L
Analog output type: output voltage signal proportional
to frequency or power
0 = frequency
1 = power
(default = 1)
2 = Pump temperature
3 = Pressure
117
R/W
N
Low Speed frequency (Hz)
1100 to “Maximum
rotational frequency”
(win 120)
(default = 1100)