Running Samples
76
NovoCyte Quanteon Flow Cytometer Operator's Guide
NOTE
The set of parameters available in the
Parameters
panel is dependent on the model
of NovoCyte Quanteon flow cytometer. The software will automatically identify the
instrument configuration and detect the available channels to be shown in
Cytometer Setting
>
Parameters
panel.
NOTE
De-selection of parameters that are not used in the experiment will help to reduce
the size of the experiment file, and thus improve the data analysis speed.
NOTE
The
Gain
column shows photodetector gains of all fluorescent channels. NovoCyte
Quanteon uses fixed photodetector gain by default. Adjustable photodetector gain
is available as advanced function. Please refer to the
NovoExpress Software Guide
for details about photodetector gain adjustment.
Set Sample Acquisition Conditions
Stop Condition
The criteria to stop sample acquisition can be set in
Cytometer Setting
>
Stop
Condition
panel (Figure 43), based on either the number of events to be collected
(all events or within a specified gate), a time limit, or a sample volume limit.
Multiple stop conditions can be selected as well. When any one of these
conditions is met, the instrument will stop collecting samples. The number of
events to be collected may be modified during the sample collection process.
However, the time or volume stop conditions cannot be modified.
Figure 43.
Set Stop Condition