Chapter 1
11
Preparing for Calibration and Performance Verification Test
Performance Verification and Adjustment Procedures
Performance Verification and Adjustment Procedures
To perform a calibration run, the performance verification tests listed in Table 1-4
below must be completed.
If any of the performance verification tests fail, perform the corresponding
calibration adjustment listed in Table 1-4.
If the performance verification tests fail and no calibration adjustment procedure is
available, the instrument must be returned to Agilent Technologies for repair. Please
contact your local Agilent Technologies Customer Sales and Service Office for
replacement parts or repair service information. See the
NFA User’s Guide
for the
address of your nearest office.
Table 1-4
Verification and adjustment listings
Test no.
Performance Verification Test
Calibration Adjustments
1
10 MHz Out Frequency Reference Accuracy
Test 2 (if applicable)
2
10 MHz Out Frequency Reference Adjustment
3
Input VSWR
4
Frequency Accuracy
5
Noise Source Supply Accuracy
6
Noise Figure Range and Accuracy
7
Gain Measurement Uncertainty
8
Instrument Noise Figure
9
Measurement Jitter
Summary of Contents for NFA Series
Page 114: ...104 Chapter3 Technical Specifications General Specifications...
Page 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Page 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Page 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Page 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Page 184: ...174 Index Index...