Agilent Technologies N9039A Installation Notes Download Page 4

4

Installation Note 

N9039-90004

Option BAB, 3.5 mm RF Connectors Option 518, Type-N RF Connectors Installation Kit

Introduction

This kit contains all of the instructions and parts required to install Option BAB or 
Option 518 on an Agilent N9039A RF Preselector.

This installation note provides instructions on how to change the RF INPUT and RF 
OUTPUT front panel connectors.

IMPORTANT

For installation of Option BAB the N9039AK-BAB license must be 
installed. 

For installation of Option BAB the license for N9039AK-BAB must be 
removed. 

Contents of N9039AK-BAB Kit

NOTE

a. Part number and memory capacity subject to change

Quantity

Description

Agilent Part 

Number

1

Installation Note

This note

1

Option Upgrade Entitlement Certificate

5964-5178

4

Screw, M3 x 0.5, 8 mm, TORX - 10

0515-0372

1

USB Storage Device 

see Note a

2

EMI O-ring shielding

8160-1637

1

Cable, semi-rigid, RF Output

N9039-21311

1

Cable, semi-rigid, RF Input

N9039-21312

1

RF Input Connector Assembly

N9039-60027

1

RF Output Connector Assembly 

N9039-60029

Summary of Contents for N9039A

Page 1: ...Installation Note Part Number N9039 90004 Printed in USA December 2008 Agilent N9039A RF Preselector Option BAB 3 5 mm RF Connectors Option 518 Type N RF Connectors...

Page 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or fo...

Page 3: ...n Kit Option BAB 3 5 mm RF Connectors Option 518 Type N RF Connectors Installation Kit Products Affected N9039A Serial Numbers All To Be Performed By X Agilent Service Center X Personnel Qualified by...

Page 4: ...tion BAB the N9039AK BAB license must be installed For installation of Option BAB the license for N9039AK BAB must be removed Contents of N9039AK BAB Kit NOTE a Part number and memory capacity subject...

Page 5: ...r N9039 90002 Personal computer with internet access and USB port T 20 TORX driver T 10 TORX driver 5 16 inch open end wrench Quantity Description Agilent Part Number 1 Installation Note This note 4 S...

Page 6: ...loth to avoid damage to the front panel keys or input connector 1 Disconnect the instrument from ac power 2 Refer to Figure 1 Using the T 20 driver remove the four screws two on each side 1 that attac...

Page 7: ...t panel are removed 5 Refer to Figure 2 Using the T 10 driver remove the eight screws 1 four on each side to detach the Front Frame Assembly from the chassis 6 Refer to Figure 3 Pull the Front Frame A...

Page 8: ...0004 Option BAB 3 5 mm RF Connectors Option 518 Type N RF Connectors Installation Kit 7 Refer to Figure 4 Loosen the three semi rigid cables from the three front panel connectors Figure 4 Front Panel...

Page 9: ...onnector 11 Remove the RF Input connector 4 and cable W3 Replace with the new RF Input connector and cable from the kit Torque the screws to 9 inch pounds and the cable to 10 inch pounds 12 Install th...

Page 10: ...ur on each side of the chassis Torque to 9 inch pounds 17 Slide the instrument cover back onto the deck from the rear The seam on the cover should be on the bottom Be sure the cover seats into the gas...

Page 11: ...ith an attached License File 4 Locate the USB Storage Device from the kit 5 Save the License File to the root directory of the USB Storage Device 6 Connect the USB Storage Device to the signal analyze...

Page 12: ...on Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C Program Files Agilent licens...

Page 13: ...e Agilent licensing manager Highlight Option BAB and press Delete Alternate Removal Procedure The License File can be manually deleted by removing the N9039A file from the following folder on the N903...

Page 14: ...er the installation of this upgrade or not If a full calibration is required arrangements regarding the level of calibration must be made between the end user and the calibration provider For assistan...

Reviews: