Agilent Technologies N5769A Service Note Download Page 2

Page 2 of 2 

 

N5769A-03 

Situation: 

When performing verification of power supply the last performance test instructs user to set the 
voltage setting to maximum rating. If power supply is set to auto restart mode it restores the operating 
setting that was set when it was last turned off. When this occurs the voltage or current will be set to 
maximum rating the next time it is turned on. This will present an unexpected voltage/current setting 
condition .   
 
 

Solution/Action: 

1.

 

During verification of instrument (s) use the latest version of users guide. 

2.

 

Download the latest user’s guide. Use web link to download latest manual found on Document 
Library tab. 

http://www.agilent.com/find/n5700

 

 
 

 

 

 

 

 

 

Revision History: 

Revision 

Number 

Date 

Author 

Reason For Change 

 

1.0 

7/11/13 

PF 

As published 

 

 

 

 

 

 