
Page 2 of 2
N5769A-03
Situation:
When performing verification of power supply the last performance test instructs user to set the
voltage setting to maximum rating. If power supply is set to auto restart mode it restores the operating
setting that was set when it was last turned off. When this occurs the voltage or current will be set to
maximum rating the next time it is turned on. This will present an unexpected voltage/current setting
condition .
Solution/Action:
1.
During verification of instrument (s) use the latest version of users guide.
2.
Download the latest user’s guide. Use web link to download latest manual found on Document
Library tab.
http://www.agilent.com/find/n5700
Revision History:
Revision
Number
Date
Author
Reason For Change
1.0
7/11/13
PF
As published