3-28
Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Performance Tests
N5242A
Performance Tests
The performance tests verify the electrical performance of your N5242A microwave PNA.
These performance tests are included in the analyzer’s firmware with Option 897 and
Option 898. Your analyzer is automatically configured for each individual test.
The model numbers of the additional equipment used by these performance tests are
specified under
“Required Service Test Equipment” on page 2-10
There are nine tests in the Option 897 or 898 performance test package:
•
•
Source Maximum Power Output Test
•
•
•
•
•
•
•
Source Power Accuracy Test
Function of the Test:
To confirm the accuracy of the source output power of your network
analyzer over its full frequency range.
Specification Tested:
Test Port Output–Power Level Accuracy
Equipment Used:
A power meter, power sensors, and adapters.
Description of the Test:
1. The analyzer’s output power level is set to 0 dBm.
2. A power sensor is connected to Port 1.
3. The output power is measured at hundreds of CW frequencies across the analyzer’s
frequency range and the values compared to the setting of 0 dBm.
If the Analyzer Fails this Test:
• Perform the
“Source Adjustment” on page 3-38
• If the analyzer still fails this test, troubleshoot the source section of the analyzer and
then repeat this test. Refer to
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...