4-16
Service Guide N5230-90014
Troubleshooting
PNA Series Microwave Network Analyzers
Front Panel Troubleshooting
N5230A
Front Panel Troubleshooting
The front panel assembly consists of the A1 keypad, A2 display, and the A3 front panel
interface. The following tests verify the operation of the front panel assembly when the
analyzer is in the measurement mode. If the instrument fails to power up correctly, or it is
difficult to verify due to a faulty display, refer to
“Power Up Troubleshooting” on page 4-6
Refer to the following sections to verify the operation of the noted assemblies.
•
“A1 Front Panel Keypad and RPG Test”
below
•
“A2 Display Test” on page 4-18
•
“A3 Front Panel Interface Board” on page 4-19
— USB
— Speaker
— Probe Power
If all assemblies are working correctly, continue troubleshooting with
.
A1 Front Panel Keypad and RPG Test
Test the front panel keypad by running the front panel test. To run the front panel test,
perform the following:
On the
System
menu, point to
Service
and click
Front Panel Test.
A
Front Panel Service Test
dialog box will be displayed, as shown in
.
Figure 4-7 Front Panel Service Test Dialog Box
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...