PNA Series Microwave Network Analyzers
Tests and Adjustments
N5230A
Non-ANSI/NCSL Z540–1–1994 Verification
Non-ANSI/NCSL Z540–1–1994 Verification
For non-ANSI/NCSL Z540-1-1994, perform the preliminary checks and the performance
tests while
stopping to troubleshoot.
Refer to
for test flow. Troubleshoot and
repair the first problem encountered without continuing to other tests. After you
troubleshoot, repair, and adjust, repeat the
last failed
portion and generate a new set of
data.
Figure 3-2
Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart