Agilent Technologies N5230-60102 Installation Note Download Page 2

2

  

                                                                                                                            

  

Installation Note N5230-90004

WARRANTY STATEMENT

THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO 
BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM 
EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, 
EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION 
CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF 
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE 
LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN 
CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR 
ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A 
SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN 
THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE 
SEPARATE AGREEMENT WILL CONTROL.

DFARS/Restricted Rights Notice

If software is for use in the performance of a U.S. Government prime contract or subcontract, 
Software is delivered and licensed as “Commercial computer software” as defined in DFAR 
252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as 
“Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent 
agency regulation or contract clause. Use, duplication or disclosure of Software is subject to 
Agilent Technologies’ standard commercial license terms, and non-DOD Departments and 
Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in 
FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than 
Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) 
(November 1995), as applicable in any technical data.

Safety Notes 

The following safety notes are used throughout this document. Familiarize yourself with each 
of these notes and its meaning before performing any of the procedures in this document. 

WARNING

Warning denotes a hazard. It calls attention to a procedure which, 
if not correctly performed or adhered to, could result in injury or 
loss of life. Do not proceed beyond a warning note until the 
indicated conditions are fully understood and met.

CAUTION

Caution denotes a hazard. It calls attention to a procedure that, if not 
correctly performed or adhered to, could result in damage to or destruction 
of the instrument. Do not proceed beyond a caution sign until the indicated 
conditions are fully understood and met.

Summary of Contents for N5230-60102

Page 1: ...Printed in USA December 2007 Supersedes July 2007 Agilent Technologies Inc 2005 2007 Installation Note Configurable Test Set Upgrade Kit Upgrade Kit Number N5230 60102 For N5230A C Standard Test Set...

Page 2: ...l item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclosure of Softwa...

Page 3: ...adding Option 014 Configurable Test Set and Option 1E1 Source Attenuators Option 014 Configurable Test Set adds the additional cabling necessary to allow your analyzer to perform measurements on high...

Page 4: ...AM to A25 port 1 step attenuator 1 N5240 20023 W41 RF cable A19 SSLAM to A26 port 2 step attenuator 1 N5240 20024 W42 RF cable A25 port 1 step attenuator to PORT 1 SOURCE OUT 1 N5240 20021 W43 RF cabl...

Page 5: ...d Static discharges too small to be felt can cause permanent damage To prevent damage to the instrument always have a grounded conductive table mat in front of your test equipment always wear a ground...

Page 6: ...nnections except the front panel cable connectors Use a 5 16 in torque wrench set to 21 in lbs for these connections Equipment Required for Post Upgrade Adjustments Description Qty Part Number T 10 TO...

Page 7: ...has not already been disconnected 2 With a T 20 TORX driver remove the strap handles item by loosening the screws item on both ends until the handle is free of the analyzer 3 With a T 20 TORX driver r...

Page 8: ...m the analyzer lift and support the front of the analyzer chassis 2 Slide the front panel over the test port connectors 3 Disconnect the front panel interface ribbon cable item The front panel is now...

Page 9: ...flat surface 2 Remove the following cables in the order listed W7 A19 SSLAM to A21 test port 1 coupler W9 A19 SSLAM R1 to A20 mixer brick R1 W11 A21 test port 1 coupler to A20 mixer brick A W8 A19 SSL...

Page 10: ...nsert each screw only a few threads into the attenuator 2 Position the attenuators in the analyzer as shown by aligning the heads of the screws with the slotted holes in the side frame Make sure the a...

Page 11: ...PORT 1 RCVR A IN to A20 mixer brick A W48 A21 test port 1 coupler to PORT 1 CPLR ARM W42 A25 port 1 step attenuator to PORT 1 SOURCE OUT W45 PORT 2 CPLR THRU to A22 test port 2 coupler W40 A19 SSLAM...

Page 12: ...12 Installation Note N5230 90004 Figure 6 New Cable Installation...

Page 13: ...scard it 3 Remove any adhesive remaining on the front panel 4 Remove the protective backing from the new front panel overlay item 5 Starting from either the left or right side loosely place the overla...

Page 14: ...using a 5 16 in torque wrench set to 21 in lbs 2 Reconnect the ribbon cable item to the A3 front panel interface board 3 Slide the front panel over the test port connectors being careful to align the...

Page 15: ...ont handles in the vertical position Do not place the analyzer on its front panel without the handles This will damage the front panel assemblies Refer to Figure 9 for this procedure 1 Slide the cover...

Page 16: ...ble 3 In the Select Desired Option list click 1E1 Source Attenuators Click Enable 4 Click Yes in answer to the displayed question in the Restart Analyzer box 5 When the installation is complete click...

Page 17: ...ound at Equipment Required for Post Upgrade Adjustments on page 6 Calibration The analyzer should now operate and phase lock over its entire frequency range If you experience difficulty with the basic...

Page 18: ...18 Installation Note N5230 90004...

Reviews: