22
Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Installation Guide
Operation Verification
Frequency Range and Accuracy Check
Spectrum Analyzer Procedure
(N5161A
1
/62A
/81A/82A/83A)
Test Setup
1.
Connect the equipment as shown.
2.
Verify that the spectrum analyzer is locked to the 10
MHz external reference frequency.
3.
Align the spectrum analyzer:
Press
System > Alignment > Align All Now
.
4.
Preset the signal generator: Press
Preset
.
5.
Turn modulation off: Press the
Mod On/Off
so that
the MOD On/Off LED turns off.
6.
Set the amplitude:
Press
Amplitude
and enter 0 dBm.
7.
Turn RF on: Press
RF On/Off
so that the RF On/Off
LED lights.
8.
Set the frequency: Press
Frequency
and set the signal
generator to the first frequency listed in
.
9.
Confirm that the measured frequency is within the
limits listed in the table.
1.For the N5161A/62A these softkey menus and features
are only available through the Web-Enabled MXG or
through SCPI commands. Refer to
Programming Guide
, and to the
SCPI Command Reference
.
10. Repeat step
for all of the frequencies in the
table that are within the frequency range of your signal
generator.
Table 3-3 Frequency Accuracy Limits
Frequency (MHz)
Limit (Hz)
N5161A/62A/8
1A/82A
Limit (Hz)
N5183A
0.1 MHz
a
a.
N5181A/82A with s/n prefix: < MY4740, measure at 0.25 MHz
±2
±2
200 MHz
±2
±2
300 MHz
±2
±2
500 MHz
±2
±2
1000 MHz
±2
±2
2000 MHz
±2
±2
3100 MHz
±2
±2
6000 MHz
±2
±4
10 GHz
--
±4
20 GHz
--
±8
32 GHz/40 GHz
--
±16