Using the Multiport Test System
How to Calibrate
Chapter 3
41
A Method of Checking Calibration
A good method of checking calibration is to establish a Golden Device, which is a device that
meets all specifications and is saved for comparison of the measured results in the future.
Establishing a Golden Device
Follow these steps to establish a golden device:
1. Calibrate the system.
2. Perform the complete set of measurements on the golden device.
3. Save and print all of the test results from these initial measurements.
Now you can measure the Golden Device when you suspect that your system may need to be
calibrated. Compare the results of these measurements against the results that you saved
and printed from the initial measurements.
Starting a Calibration
1. Complete the hardware and software setup. See
“How to Set Up the Multiport Test System
and
“How to Install and Start the Balanced Measurement
.
2. Define a test plan to include desired measurement types, and the use and placement of any
adapters.
3. Characterize the adapters. See
“How to Characterize Adapters” on page 44
.
4. For test sets with the optional step attenuator, determine the attenuation needs. Once you
calibrate at an attenuation setting, any changes to the setting will invalidate the accuracy
of the calibration. Even if the attenuator is moved from the calibrated position and reset
prior to measurement, measurement accuracy will be reduced. It is strongly recommended
that you recalibrate if you have changed the attenuator setting.
5. When all of these prerequisite steps have been completed, select
Calibrate > Standard
Calibration…
or
Calibrate > Electronic Calibration…
to begin. You are guided
through this process step-by-step.
NOTE
Creating a Calibration Subset
If you do not want to use all frequency points for your measurement, you can
create a calibration subset and use it for the active calibration instead of the
full calibration set. See
“Calibrate > Create Cal Subset…” on page 133
.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...