Maintenance and Troubleshooting
Connector Care for RF & Microwave Coaxial Connectors
Chapter 6
158
Connector Care Quick Reference
Handling and Storage
Do
Keep connectors clean
Do Not
Touch mating-plane surfaces
Extend sleeve or connector nut
Set connectors contact-end down
Use plastic end-caps during storage
Visual Inspection
Do
Inspect all connectors carefully
Do Not
Use a damaged connector - ever
Look for metal particles, scratches,
and dents
Connector Cleaning
Do
Try compressed air first
Do Not
Use any abrasives
Use isopropyl alcohol
Get liquid into plastic support
beads
Clean connector threads
Gaging Connectors
Do
Clean and zero the gage before use
Do Not
Use an out-of-spec connector
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Making Connections
Do
Align connectors carefully
Do Not
Apply bending force to connection
Make preliminary connection lightly
Over tighten preliminary
connection
Turn only the connector nut
Twist or screw any connection
Use a torque wrench for final
connect
Tighten past torque wrench
"break" point
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...