Instructions for Use
Installation
4
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (E)
43/90
Installation
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
temperature: from +12 °C to +40 °C
relative humidity: 0 – 95 % (non-condensing)
IP value.
CAUTION!
Before starting the pump, check the oil level.
WARNING!
Take out the protective caps on the suction and exhaust flanges before doing
anything else. In the event of an accidental start-up, the air inside the pump
could violently expel the protective caps and harm the operator.
WARNING!
During installation, pay maximum attention that the suction flange is
connected to the vacuum chamber and the exhaust flange is not closed (see
the following figure). The pump must not be used as a compressor. Maximum
pressure inside the oil container must not exceed 1.5 bar (abs.). Non-
observance of these precautions may be dangerous for the machine and the
operator.