Agilent Technologies K6460 Safety Manual Download Page 4

Safety

Symbols on Instruments

4

K6460 Triple Quad MS Safety Guide

Do not operate the instrument in the presence of 
hydrogen or any other flammable gases or fumes. 
Operation of any electrical instrument in such an 
environment constitutes a definite safety hazard.

Ne pas utiliser l'instrument en présence de gaz ou de 
vapeurs inflammables. L'utilisation de tout 
instrument électrique dans un tel environnement 
constitue un danger certain.

Do not discard this electrical/electronic product in 
domestic household waste.

Ne jetez pas ce produit électrique / électronique 
avec les ordures ménagères.

Disconnect main plug from electrical outlet prior to 
fuse replacement.

Débranchez la prise principale de la prise électrique 
avant de remplacer le fusible.

Two live circuits. Disconnect both power cords 
before maintenance.

Deux circuits sous tension. Débranchez les deux 
cordons d'alimentation avant l'entretien.

Biological risks

Dangers biologiques

Table 2

Safety symbols (continued)

Symbol

Description

Summary of Contents for K6460

Page 1: ...formity and Standards 12 Safety and Regulatory Certifications 12 Electromagnetic Compatibility 13 Environmental Conditions 14 Sound Emission Declaration 14 This guide contains information about safety...

Page 2: ...s The following table describes the electrical symbols that can be displayed on the K6460 mass spectrometer Table 1 Electrical symbols Symbol Description Indicates a protective earth terminal Indique...

Page 3: ...operator and to protect the apparatus against damage L appareil est marqu avec ce symbole lorsque l utilisateur doit se r f rer au manuel d instruction afin de prot ger contre les atteintes l op rateu...

Page 4: ...l environnement constitue un danger certain Do not discard this electrical electronic product in domestic household waste Ne jetez pas ce produit lectrique lectronique avec les ordures m nag res Disco...

Page 5: ...Read and understood all material safety data sheets MSDS for chemicals handled WARNING PHYSICAL INJURY HAZARD Use the K6460 mass spectrometer only as specified in the user instructions Using the syst...

Page 6: ...r cord and the AC power supply In the AC power supply In the wiring from the AC power supply to the power switch WARNING Never remove a cover unless specifically instructed to do so With the power swi...

Page 7: ...s with adequate current capacity The branch circuit over current protection Circuit Breakers protecting the building receptacles must be rated no more than 20A WARNING Be sure that the LC MS main supp...

Page 8: ...s equipment must be installed in a Category II environment as defined in IEC 664 For all instruments the supply voltage must not fluctuate more that 10 For any instrument with an Agilent Jet Stream so...

Page 9: ...es or protective clothing Minimize the inhalation of chemicals Do not leave chemical containers open Use only with adequate ventilation for example fume hood Check regularly for chemical leaks or spil...

Page 10: ...n a fume hood After emptying a waste container seal it with the cap provided Dispose of the contents of the waste tray and waste bottle in accordance with good laboratory practices and local state pro...

Page 11: ...r aerosol protection Always follow local state and federal biohazard handling regulation when disposing of biohazardous waste material Handle all infectious samples according to good laboratory proced...

Page 12: ...gies the protections provided by the instrument may be impaired Safety and Regulatory Certifications The Agilent 6460 Triple Quad Mass Spectrometer Model K6460 conforms to applicable editions and vers...

Page 13: ...off and on try one or more of the following measures 1 Relocate the susceptible equipment 2 Move the mass spectrometer away from the susceptible equipment 3 Plug the devices into different electrical...

Page 14: ...kpegel Lp 70 dB Environmental Conditions Equipment Class Class I Laboratory Equipment Pollution Degree 2 Installation Category II Environment Indoor Use Altitude Not to exceed 3000 m Electrical supply...

Page 15: ...This page intentionally left blank...

Page 16: ...ry information for your Agilent 6460 Triple Quad Mass Spectrometer Model K6460 system For In Vitro Diagnostic Use The K6460 mass spectrometer is intended to be used to identify inorganic or organic co...

Reviews: