How to Prepare the Eye Opening Measurement
To prepare an Eye Opening measurement to test a shielded cable:
1
Disable the pattern generator outputs by pressing the
0V (Disable)
button in
the
PG Setup -> Data Output
screen.
2
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
3
Terminate all non-connected pattern generator output ports with 50 Ω.
4
Switch to the
Pattern
panel and press
Pattern Select
. Select an appropriate
pattern, for example a pure 2^15-1 PRBS, for both the pattern generator and
the error detector.
5
For the pattern generator setup you need to specify the logic levels and the bit
rate. Select
ECL
levels and a clock speed of 1250 MHz in this example. This
corresponds to a clock period of 0.8ns. See
“Setting up the Pattern Generator
for more information.
6
Set up the error detector so that the input range and the termination matches
the pattern generator's levels:
– Select an
Input Range
from -2V to 0V
– Set the
Data Termination
to -2V
– Set the
Alignment BER Threshold
to 1E-6.
– Set the
Clock Setup
to
Clock Data Recovery
to get the error detector's
clock from the incoming data stream.
7
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
8
Press
Sync Now
and then
Auto Align
to find the optimum sampling point. Check
that the synchronization and the alignment were successful. None of the error
indicators should show red.
The resulting BER should be zero.
6
Advanced Analysis
290
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...